PHILOS DR US
Report
- Report Number
- 1028232-2010-02716
- Event Type
- Malfunction
- Date Received
- April 29, 2011
- Date of Event
- December 8, 2009
- Report Date
- December 8, 2009
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- NVZ
- PMA / PMN Number
- P950037
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- NURSE
Narratives
UPON RECEIPT, THE PACEMAKER WAS SUBJECTED TO AN ELECTRICAL INCOMING INSPECTION. THE BATTERY OF THE PACEMAKER WAS FOUND TO BE FULLY DEPLETED. THEREFORE, THE PACEMAKER'S PACING SIGNAL COULD NOT BE OBSERVED. A DESTRUCTIVE ANALYSIS CONFIRMED A FULLY DEPLETED BATTERY. CONNECTING AN EXTERNAL POWER SUPPLY, THE PACEMAKER SHOWED EXPECTED PACING AND SENSING SIGNALS. THERE WAS NO INDICATION OF SENDING AND/OR PACING PROBLEMS. THE ANALYSIS OF THE CURRENT CONSUMPTION SHOWED EXPECTED VALUES. ALSO THE CIRCUIT WAS FUNCTIONED CORRECTLY. THE COMPLAINT DESCRIPTION SHOWS AN ESTIMATED REMAINING SERVICE TIME WITH UNEXPECTED VALUES. HOWEVER, DESPITE OF THOROUGH ANALYSIS THE CLINICAL BEHAVIOR COULD NOT BE REPRODUCED. THE ESTIMATED SERVICE TIME FOR THIS PACEMAKER IS 66 MONTHS TO ERI FOR FACTORY SETTINGS WITH 100% STIMULATION. THE OBSERVED SERVICE TIME LIES WITH 115 MONTHS WELL ABOVE THE EXPECTED RANGE. IN SUMMARY, THE BATTERY PROVED TO BE DEPLETED. THE ANALYSIS CONFIRMED THAT THE DISPLAYED REMAINING SERVICE TIME WAS INACCURATE. DURING ANALYSIS THE CLINICAL BEHAVIOR WAS NOT REPRODUCIBLE.
THIS DEVICE SHOWED 6 YRS 10 MONTHS REMAINING UNTIL ERI. BATTERY IMPEDANCE WAS 10 KOHMS AND VOLTAGE WAS 2.41. IT WAS RECOMMENDED TO EXPLANT AT THAT TIME AND SEND THE DEVICE IN FOR ANALYSIS. ON (B)(6) 2010 - THIS DEVICE WAS RETURNED AND IS AT ERI INDICATION AND HAS MET LONGEVITY ESTIMATES. THIS WAS EXPLANTED ON (B)(6) 2010 AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PHILOS DR US | PACEMAKER | NVZ | BIOTRONIK SE & CO. KG | 331443 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Hospitalization |