FDA Adverse Event Malfunction Summary report: N

CONTAK RENEWAL

MDR report key: 2161585 · Received July 14, 2011

Report

Report Number
2124215-2011-08168
Event Type
Malfunction
Date Received
July 14, 2011
Date of Event
April 18, 2011
Report Date
December 18, 2012
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
PMA / PMN Number
P010012
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MT
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. PIN GAUGE TESTING, DESIGNED TO VERIFY PROPER PORT DIMENSIONS, WAS COMPLETED. EACH PORT MEASURED AS EXPECTED. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED. THE CLINICAL OBSERVATIONS WERE NOT ABLE TO BE CONFIRMED.

Additional Manufacturer Narrative · 1

AS OF TODAY, NO ADDITIONAL INFORMATION IS AVAILABLE AND AT THIS TIME, OUR INVESTIGATION IS COMPLETE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

SUBSEQUENT INFORMATION INDICATED THAT THE DEVICE WAS EXPLANTED FOR NORMAL BATTERY DEPLETION. THE DEVICE WAS RETURNED FOR ANALYSIS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR LEAD AND CARDIAC RESYNCHRONIZATION THERAPY-DEFIBRILLATOR (CRT-D) DISPLAYED A SINGLE EPISODE OF NOISE. THE NOISE WAS OVERSENSED LEADING TO PACING INHIBITION WITH ASYSTOLE. NOISE WAS NOT ABLE TO BE RECREATED. ALL LEAD DIAGNOSTICS WERE NORMAL. THE PHYSICIAN PLANED TO MONITOR THE SYSTEM FOR THE TIME BEING. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED. THE SYSTEM REMAINS IN SERVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H175

Patients

Seq Age Sex Outcome Treatment
1 87 YR 4524| H175| 4244| 0158