FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2161574 · Received July 14, 2011

Report

Report Number
2531779-2011-04934
Event Type
Malfunction
Date Received
July 14, 2011
Date of Event
April 19, 2011
Report Date
April 19, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS MADE BASED ON THE RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2011. CORRECTION/REMOVAL NUMBER: 2531779-03/24/2010-003-R. DURING EVALUATION THE FORCE SENSOR ASSEMBLY WAS FOUND TO BE PHYSICALLY DAMAGED.

Description of Event or Problem · 1

PUMP IS RETURNED FOR MULTIPLE LOSS OF PRIME ALARMS. EVALUATION REVEALED A DAMAGED FORCE SENSOR ASSEMBLY. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.