CONTAK RENEWAL
Report
- Report Number
- 2124215-2011-08100
- Event Type
- Injury
- Date Received
- July 14, 2011
- Date of Event
- April 15, 2011
- Report Date
- June 17, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NIK
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, TECHNICIANS USED AN ENGINEERING-LEVEL LONGEVITY PREDICTION CALCULATION TO ASSESS THE RATE OF BATTERY DEPLETION, GIVEN THE PROGRAMMED PARAMETERS AND OTHER DATA STORED WITHIN THE MEMORY OF THE DEVICE. THE RESULTS OF THIS CALCULATION INDICATED THAT THE ACTUAL RATE OF BATTERY DEPLETION FELL WITHIN AN ACCEPTABLE RANGE. NEXT, A COMPREHENSIVE SERIES OF DIAGNOSTIC TESTS WERE CONDUCTED THAT VERIFIED THE PERFORMANCE OF THERAPY DELIVERY, SENSING, AND RECORDING FUNCTIONS. HAVING MET THE ENGINEERING LONGEVITY PREDICTION, FUNCTIONALLY PASSING ALL RETURNED PRODUCT TESTING, AND WITH NO FURTHER INFORMATION TO INDICATE A PRODUCT PERFORMANCE ISSUE, WE HAVE CONCLUDED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION.
THE DEVICE IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE HAD REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) 56 MONTHS POST IMPLANT. IT WAS ALLEGED THE DEVICE HAD DEPLETED PREMATURELY. A REPLACEMENT SURGERY WAS PERFORMED AND THE DEVICE WAS EXPLANTED. NO FURTHER PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTAK RENEWAL | IMPLANTABLE CHF GENERATOR | NIK | GUIDANT CRM CLONMEL IRELAND | H195 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |