ION PACLITAXEL-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM
Report
- Report Number
- 2134265-2011-02656
- Event Type
- Injury
- Date Received
- July 14, 2011
- Date of Event
- June 14, 2011
- Report Date
- June 14, 2011
- Manufacturer
- BOSTON SCIENTIFIC - MAPLE GROVE
- Product Code
- NIQ
- PMA / PMN Number
- P100023
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE IS A COMBINATION PRODUCT. DEVICE EVALUATED BY MFR: THE DEVICE HAS NOT BEEN RECEIVED FOR ANALYSIS. UPON RECEIPT AND COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED. (B)(4).
IT WAS REPORTED THAT DURING A CORONARY ARTERY STENTING TREATMENT PROCEDURE A STENT DISLODGEMENT OCCURRED. THE LESION BEING TREATED WAS A 99% STENOSIS OF THE HEAVILY CALCIFIED RIGHT CORONARY ARTERY (RCA). THE LESION WAS APPROACHED VIA THE RIGHT RADIAL ARTERY. FOLLOWING DIAGNOSTIC VIEWS OF BOTH THE LEFT AND RIGHT CORONARY SYSTEMS, A 6F JR4 GUIDE CATHETER WAS ADVANCED AND A 0.014" NON-BSC GUIDE WIRE WAS ADVANCED ACROSS THE LESION. THE PHYSICIAN INTENDED TO DIRECT STENT THE RCA LESION USING A 3.5X12MM ION STENT; HOWEVER, THE STENT CAME OFF THE BALLOON IN THE RCA. THE PHYSICIAN ATTEMPTED TO SNARE THE ION STENT USING A 1.5X8MM APEX BALLOON, BUT WAS UNSUCCESSFUL. A NON-BSC SNARE WAS INTRODUCED AND RE-CAPTURED THE ION STENT. AT THAT POINT, THE RADIAL APPROACH WAS DISCONTINUED AND THE RIGHT FEMORAL WAS ACCESSED FOR FURTHER INTERVENTION. ADDITIONAL ATTEMPTS AT BALLOON ANGIOPLASTY USING A 2.5X12MM APEX AND 2.25X12MM NON-BSC BALLOON WERE UNSUCCESSFUL DUE TO INABILITY TO CROSS THE LESION. THE PATIENT WAS TRANSFERRED TO ANOTHER FACILITY FOR FURTHER INTERVENTION. THE PATIENT STATUS WAS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ION PACLITAXEL-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM | CORONARY DRUG-ELUTING STENT | NIQ | BOSTON SCIENTIFIC - MAPLE GROVE | H7493902412350 | 14020654 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | Required Intervention |