- Report Number
- 6000001-2011-11783
- Event Type
- Malfunction
- Date Received
- July 13, 2011
- Date of Event
- May 31, 2011
- Report Date
- June 17, 2011
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K915522
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PR
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A FLO-GARD INFUSION PUMP WHICH ALARMED "OCCLUSION ALL TIME" WAS CONFIRMED DURING PRODUCT EVALUATION. THIS CONDITION WAS CAUSED BY THE DOOR BEING BROKEN IN THE LATCH AREA. THE PUMP HEAD DOOR WAS REPLACED TO CORRECT THE REPORTED CONDITION. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. (B)(4). BAXTER WILL CONTINUE TO MONITOR AND REPORT ON DEATH AND SERIOUS INJURY EVENTS AND FOLLOW EXISTING ESCALATION PROCESSES TO ASSESS THE IMPACT ON PATIENT SAFETY.
(B)(4). A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE FACILITY REPRESENTATIVE CONTACTED BAXTER TO REPORT A FLOGARD PUMP IN WHICH "OCCLUSION ALL TIME" OCCURRED. THERE WAS NO PATIENT INVOLVEMENT. THERE WAS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |