FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2161403 · Received July 13, 2011

Report

Report Number
2531779-2011-04922
Event Type
Malfunction
Date Received
July 13, 2011
Date of Event
April 17, 2011
Report Date
April 17, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(6). CORRECTION NUMBER: 2531779-03/24/2010/003-R. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2011. DEVICE EVALUATION: A REVIEW OF THE PUMP HISTORY DID NOT SHOW ANY ALARMS RELATED TO THE EVENT. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO LOSS OF PRIME WARNINGS BEING DUPLICATED. EVALUATION REVEALED THAT THE FORCE SENSOR ASSEMBLY WAS DAMAGED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR MULTIPLE LOSS OF PRIME WARNINGS. EVALUATION REVEALED A PHYSICALLY DAMAGED FORCE SENSOR ASSEMBLY. THIS REPORT IS BEING MADE BASED ON RESULTS OF EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 15