FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2161403
·
Received July 13, 2011
Report
- Report Number
- 2531779-2011-04922
- Event Type
- Malfunction
- Date Received
- July 13, 2011
- Date of Event
- April 17, 2011
- Report Date
- April 17, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
(B)(6). CORRECTION NUMBER: 2531779-03/24/2010/003-R. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2011. DEVICE EVALUATION: A REVIEW OF THE PUMP HISTORY DID NOT SHOW ANY ALARMS RELATED TO THE EVENT. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO LOSS OF PRIME WARNINGS BEING DUPLICATED. EVALUATION REVEALED THAT THE FORCE SENSOR ASSEMBLY WAS DAMAGED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
Description of Event or Problem · 1
THE PUMP WAS RETURNED FOR MULTIPLE LOSS OF PRIME WARNINGS. EVALUATION REVEALED A PHYSICALLY DAMAGED FORCE SENSOR ASSEMBLY. THIS REPORT IS BEING MADE BASED ON RESULTS OF EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 |