FDA Adverse Event
Malfunction
Summary report: N
HOMECHOICE AUTOMATED PD SET WITH CASSETTE
MDR report key: 2161396
·
Received July 13, 2011
Report
- Report Number
- 1423500-2011-09020
- Event Type
- Malfunction
- Date Received
- July 13, 2011
- Date of Event
- May 28, 2011
- Report Date
- June 23, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). A FOLLOW UP REPORT WILL BE FILED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
Additional Manufacturer Narrative · 1
(B)(4). THE ROOT CAUSE OF THE SYSTEM ERROR 2240 WAS UNDETERMINED. THE LOT NUMBER IS UNKNOWN THEREFORE A BATCH REVIEW WAS NOT PERFORMED. BAXTER HAS FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH RENQ-CAPA-(B)(4).
Description of Event or Problem · 1
A SYSTEM ERROR (SE) 2240 ALARM WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE DEVICE. A SE 2240 IS A NON-RECOVERABLE ALARM THAT OCCURS WHEN THE DEVICE HAS DETECTED AIR BEING DRAWN CONTINUOUSLY INTO THE DISPOSABLE. IT IS UNKNOWN IF THIS ALARM OCCURRED DURING PATIENT THERAPY, HOWEVER, NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.