ION PACLITAXEL-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM
Report
- Report Number
- 2134265-2011-02726
- Event Type
- Malfunction
- Date Received
- July 13, 2011
- Date of Event
- May 25, 2011
- Report Date
- June 22, 2011
- Manufacturer
- BOSTON SCIENTIFIC - MAPLE GROVE
- Product Code
- NIQ
- PMA / PMN Number
- P100023
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
AGE AT TIME OF EVENT: 18 YEARS OR OLDER. (B)(4). DEVICE IS A COMBINATION PRODUCT DEVICE EVALUATED BY MFR: RETURNED PRODUCT CONSISTED OF A TAXUS ELEMENT/ION MR STENT DELIVERY SYSTEM (SDS). BLOOD AND CONTRAST WERE IN THE DISTAL AND MID SHAFT OF THE DEVICE WHICH IS CONSISTENT WITH THE REPORTED INFORMATION THAT THE DEVICE WAS USED. IT WAS DETERMINED THAT ROWS THREE THROUGH FIVE ON THE DISTAL END OF THE STENT HAD STRUTS THAT WERE STRETCHED AND FLARED. THE DISTAL TIP OF THE CATHETER WAS DAMAGED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).
REPORTABLE BASED ON ANALYSIS COMPLETED ON (B)(6) 2011. IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY TREATMENT PROCEDURE DIFFICULTY CROSSING OCCURRED. ACCESS WAS GAINED THROUGH THE FEMORAL ARTERY. THE "MODERATELY" STENOSED LESION BEING TREATED WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. THE 2.75X12MM ION MONORAIL DRUG ELUTING STENT DELIVERY SYSTEM WAS ADVANCED TO THE TARGET LESION, BUT WAS UNABLE TO CROSS THE LESION. THE PROCEDURE WAS COMPLETED WITH ANOTHER ION STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT STATUS IS FINE. UPON ANALYSIS OF THE RETURNED DEVICE, STENT DAMAGE WAS FOUND.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ION PACLITAXEL-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM | CORONARY DRUG-ELUTING STENT | NIQ | BOSTON SCIENTIFIC - MAPLE GROVE | H7493902412270 | 13990132 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |