FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2161337 · Received July 13, 2011

Report

Report Number
6000001-2011-11742
Event Type
Malfunction
Date Received
July 13, 2011
Date of Event
June 28, 2011
Report Date
June 29, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Removal / Correction Number
6000001-08/04/10-002-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON SITE BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY ALARM WAS CONFIRMED AND DUPLICATED BY BAXTER SERVICE PERSONNEL. THE ROOT CAUSE OF THIS CONDITION WAS DETERMINED TO BE DEPLETED MAIN BATTERIES AS A RESULT OF USER ERROR. THE MAIN BATTERIES AND HARNESS WERE REPLACED TO CORRECT THIS CONDITION. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.13.90. ADDITIONAL INFORMATION: A DEVICE HISTORY REVIEW REVEALED NO ABNORMALITIES WERE FOUND DURING MANUFACTURING. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY ALARM. IT IS UNKNOWN WHEN, OR IN WHICH CARE AREA, THIS EVENT OCCURRED. THIS CONDITION HAS THE POTENTIAL TO INTERRUPT DELIVERY. THE FACILITY REPRESENTATIVE STATED THAT THERE WAS NO PATIENT INVOLVEMENT, PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION IS UNKNOWN AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1