FDA Adverse Event Malfunction Summary report: N

CD HORIZON SPINAL SYSTEM

MDR report key: 21613315 · Received March 15, 2025

Report

Report Number
1030489-2025-01384
Event Type
Malfunction
Date Received
March 15, 2025
Date of Event
January 24, 2024
Report Date
April 9, 2025
Manufacturer
MDT SOFAMOR DANEK PUERTO RICO MFG
Product Code
KWP
UDI-DI
00885074457093
PMA / PMN Number
K052187
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

G2: COUNTRY OF ORIGIN IS CANADA. PRODUCT ANALYSIS OF PART # (B)(4), LOT # H5863375 VISUAL INSPECTION REVEALED THE STARTING PORTION OF THE THREADS HAS BEEN STRIPPED. DAMAGE PRESENT IS CONSISTENT WITH TORSIONAL OVERLOAD. H11: THIS REGULATORY REPORT IS BEING SUBMITTED DUE TO RETROSPECTIVE REVIEW THROUGH CAPA 624392. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

G2: COUNTRY OF ORIGIN IS CHINA. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA MANUFACTURER REPRESENTATIVE REGARDING SPINAL PRODUCT THAT WILL BE USED IN POSTERIOR LUMBAR INTERBODY FUSION SPINAL CANAL DECOMPRESSION SITE FOR LUMBAR SPINAL STENOSIS. IT WAS REPORTED THAT DURING THE IMPLANTATION PROCESS OF THE SET SCREW, THE THREAD WAS SLIPPERY DUE TO DEFORMATION AND THE LOCKING COULD NOT BE COMPLETED, SO THE IMPLANTATION PROCESS WAS LIMITED. THERE WERE NO FURTHER COMPLICATIONS OR SYMPTOMS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
547236 CD HORIZON SPINAL SYSTEM APPLIANCE, FIXATION, SPINAL INTERLAMINAL KWP MDT SOFAMOR DANEK PUERTO RICO MFG 7540020 H5863375 00885074457093

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown