CD HORIZON SPINAL SYSTEM
Report
- Report Number
- 1030489-2025-01384
- Event Type
- Malfunction
- Date Received
- March 15, 2025
- Date of Event
- January 24, 2024
- Report Date
- April 9, 2025
- Manufacturer
- MDT SOFAMOR DANEK PUERTO RICO MFG
- Product Code
- KWP
- UDI-DI
- 00885074457093
- PMA / PMN Number
- K052187
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
G2: COUNTRY OF ORIGIN IS CANADA. PRODUCT ANALYSIS OF PART # (B)(4), LOT # H5863375 VISUAL INSPECTION REVEALED THE STARTING PORTION OF THE THREADS HAS BEEN STRIPPED. DAMAGE PRESENT IS CONSISTENT WITH TORSIONAL OVERLOAD. H11: THIS REGULATORY REPORT IS BEING SUBMITTED DUE TO RETROSPECTIVE REVIEW THROUGH CAPA 624392. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
G2: COUNTRY OF ORIGIN IS CHINA. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA MANUFACTURER REPRESENTATIVE REGARDING SPINAL PRODUCT THAT WILL BE USED IN POSTERIOR LUMBAR INTERBODY FUSION SPINAL CANAL DECOMPRESSION SITE FOR LUMBAR SPINAL STENOSIS. IT WAS REPORTED THAT DURING THE IMPLANTATION PROCESS OF THE SET SCREW, THE THREAD WAS SLIPPERY DUE TO DEFORMATION AND THE LOCKING COULD NOT BE COMPLETED, SO THE IMPLANTATION PROCESS WAS LIMITED. THERE WERE NO FURTHER COMPLICATIONS OR SYMPTOMS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 547236 | CD HORIZON SPINAL SYSTEM | APPLIANCE, FIXATION, SPINAL INTERLAMINAL | KWP | MDT SOFAMOR DANEK PUERTO RICO MFG | 7540020 | H5863375 | 00885074457093 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |