FDA Adverse Event Malfunction Summary report: N

ORTHO SUMMIT SAMPLE HANDLING SYSTEM

MDR report key: 2161322 · Received July 13, 2011

Report

Report Number
2250051-2011-00166
Event Type
Malfunction
Date Received
July 13, 2011
Date of Event
July 8, 2011
Report Date
July 12, 2011
Manufacturer
HAMILTON BONADUZ AG
Product Code
MMH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT HAS BEEN INVESTIGATED. THE FIELD SERVICE ENGINEER (FSE) INSPECTED THE INSTRUMENT. THE FSE FOUND THE PLUNGER CLAMP IN POSITION #8 WAS FAILING PULL FORCE. THE FSE REPLACED THE CLAMP AND LLD SPRING AND VERIFIED THE HOLD AND PULL FORCE ON ALL THE OTHER POSITIONS. THE FSE PERFORMED SPLLD CHECKING PROCEDURE AND TESTED SUMMIT WITH OAS USER SOFTWARE. THE INSTRUMENT WAS TESTED WITHOUT PROBLEM AND WAS RETURNED TO EXPECTED OPERATION.

Description of Event or Problem · 1

THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT REPORTEDLY DID NOT PIPETTE SAMPLE AND REAGENT AND DID NOT GENERATE AN ERROR MESSAGE. NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ORTHO SUMMIT SAMPLE HANDLING SYSTEM SAMPLE PIPETTER MMH HAMILTON BONADUZ AG 936400

Patients

Seq Age Sex Outcome Treatment
1