INSIGNIA
Report
- Report Number
- 2124215-2011-06835
- Event Type
- Malfunction
- Date Received
- July 13, 2011
- Date of Event
- April 6, 2011
- Report Date
- February 28, 2012
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NVZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
A BOSTON SCIENTIFIC TECHNICAL SERVICES CONSULTANT DOCUMENTED THE REPORTED CLINICAL OBSERVATION. THE DEVICE REMAINS IMPLANTED AND NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS ISSUE WILL BE RE-EVALUATED IF ADDITIONAL INFORMATION IS RECEIVED.
UPON RECEIPT AT (B)(4) LABORATORY, AN ENGINEERING-LEVEL LONGEVITY PREDICTION CALCULATION WAS COMPLETED TO ASSESS THE RATE OF BATTERY DEPLETION. GIVEN THE PROGRAMMED PARAMETERS AND OTHER DATA STORED WITHIN THE MEMORY OF THE DEVICE, THE RESULTS OF THIS CALCULATION INDICATED THAT THE ACTUAL RATE OF BATTERY DEPLETION FELL WITHIN AN ACCEPTABLE RANGE. NEXT, A SERIES OF DIAGNOSTIC TESTS WERE CONDUCTED THAT VERIFIED THE PERFORMANCE OF PACING, SENSING, AND RECORDING FUNCTIONS. HAVING MET THE ENGINEERING LONGEVITY PREDICTION, FUNCTIONALLY PASSING ALL RETURNED PRODUCT TESTING, AND WITH NO FURTHER INFORMATION TO INDICATE A PRODUCT PERFORMANCE ISSUE, WE HAVE CONCLUDED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION.
THE DEVICE WAS SUBSEQUENTLY EXPLANTED FOR NORMAL BATTERY DEPLETION AND WAS RETURNED FOR TESTING. THIS PRODUCT ISSUE WILL BE UPDATED WHEN DEVICE EVALUATION IS COMPLETE.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE IS DISPLAYING A LONGEVITY REMAINING OF (B)(6). HOWEVER, THE MAGNET RATE IS 90PPM WHICH CORRESPONDS WITH THE INDICATOR OF ELECTIVE REPLACEMENT NEAR (ERN).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INSIGNIA | IMPLANTABLE PULSE GENERATOR | NVZ | GUIDANT CRM CLONMEL IRELAND | 1298 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR | Hospitalization| L| R | 1298| 4469| 4456 |