FDA Adverse Event Malfunction Summary report: N

INSIGNIA

MDR report key: 2161256 · Received July 13, 2011

Report

Report Number
2124215-2011-06835
Event Type
Malfunction
Date Received
July 13, 2011
Date of Event
April 6, 2011
Report Date
February 28, 2012
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NVZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

A BOSTON SCIENTIFIC TECHNICAL SERVICES CONSULTANT DOCUMENTED THE REPORTED CLINICAL OBSERVATION. THE DEVICE REMAINS IMPLANTED AND NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS ISSUE WILL BE RE-EVALUATED IF ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT (B)(4) LABORATORY, AN ENGINEERING-LEVEL LONGEVITY PREDICTION CALCULATION WAS COMPLETED TO ASSESS THE RATE OF BATTERY DEPLETION. GIVEN THE PROGRAMMED PARAMETERS AND OTHER DATA STORED WITHIN THE MEMORY OF THE DEVICE, THE RESULTS OF THIS CALCULATION INDICATED THAT THE ACTUAL RATE OF BATTERY DEPLETION FELL WITHIN AN ACCEPTABLE RANGE. NEXT, A SERIES OF DIAGNOSTIC TESTS WERE CONDUCTED THAT VERIFIED THE PERFORMANCE OF PACING, SENSING, AND RECORDING FUNCTIONS. HAVING MET THE ENGINEERING LONGEVITY PREDICTION, FUNCTIONALLY PASSING ALL RETURNED PRODUCT TESTING, AND WITH NO FURTHER INFORMATION TO INDICATE A PRODUCT PERFORMANCE ISSUE, WE HAVE CONCLUDED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION.

Description of Event or Problem · 1

THE DEVICE WAS SUBSEQUENTLY EXPLANTED FOR NORMAL BATTERY DEPLETION AND WAS RETURNED FOR TESTING. THIS PRODUCT ISSUE WILL BE UPDATED WHEN DEVICE EVALUATION IS COMPLETE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE IS DISPLAYING A LONGEVITY REMAINING OF (B)(6). HOWEVER, THE MAGNET RATE IS 90PPM WHICH CORRESPONDS WITH THE INDICATOR OF ELECTIVE REPLACEMENT NEAR (ERN).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INSIGNIA IMPLANTABLE PULSE GENERATOR NVZ GUIDANT CRM CLONMEL IRELAND 1298

Patients

Seq Age Sex Outcome Treatment
1 81 YR Hospitalization| L| R 1298| 4469| 4456