FINELINE II
Report
- Report Number
- 2124215-2011-08436
- Event Type
- Malfunction
- Date Received
- July 13, 2011
- Date of Event
- April 14, 2011
- Report Date
- April 14, 2011
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVN
- PMA / PMN Number
- P960004
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE LEAD WAS NOT RETURNED FOR TESTING. THEREFORE, BOSTON SCIENTIFIC CANNOT CONFIRM THE REPORTED CLINICAL OBSERVATIONS. NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS ISSUE WILL BE RE-EVALUATED IF ADDITIONAL INFORMATION IS RECEIVED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS ATRIAL LEAD WAS EXHIBITING NOISE AND BIPOLAR IMPEDANCE MEASUREMENTS GREATER THAN 2500 OHMS. THE OUT OF RANGE IMPEDANCE MEASUREMENTS TRIGGERED THE LEAD SAFETY SWITCH (LSS) ON THE ASSOCIATED PACEMAKER. UNIPOLAR IMPEDANCE MEASUREMENTS WERE 500 OHMS AND THE LEAD WAS PROGRAMMED TO UNIPOLAR CONFIGURATION. ADDITIONALLY, NO SENSING MEASUREMENTS WERE OBSERVED FROM THIS LEAD WITH THE PACING SYSTEM ANALYZER (PSA). A REVISION PROCEDURE WAS PERFORMED AND THE LEAD WAS REMOVED FROM SERVICE AND REPLACED WITHOUT FURTHER INCIDENT. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FINELINE II | IMPLANTABLE LEAD | NVN | CPI - DEL CARIBE | 4469 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR | (B)(4)| (B)(4)| (B)(4) |