FDA Adverse Event Malfunction Summary report: N

INFANT CONTINUOUS FLOW BREATHING CIRCUIT

MDR report key: 2161131 · Received July 13, 2011

Report

Report Number
9611451-2011-00419
Event Type
Malfunction
Date Received
July 13, 2011
Date of Event
March 29, 2011
Report Date
June 23, 2011
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K020332
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT DEVICE WAS RECEIVED IN AN UNSEALED BAG AND WAS VISUALLY INSPECTED. RESULTS: IT WAS OBSERVED THAT THE OFFSET ADAPTOR WAS MISSING FROM THE END OF THE INSPIRATORY LIMB. A LOT CHECK REVEALED ONE OTHER COMPLAINT FOR THIS LOT NUMBER. CONCLUSION: THE LIKELY CAUSE OF THIS IS FAILURE OF PRODUCTION LINE STAFF TO CONNECT THIS COMPONENT WHEN ASSEMBLING THE BREATHING CIRCUIT AND FAILURE OF PRODUCTION LINE STAFF TO CHECK THIS COMPONENT DURING VISUAL INSPECTION OF THE BREATHING CIRCUIT. OPERATING PROCEDURES ARE IN PLACE TO ASSIST THE OPERATORS ON THE PRODUCTION LINE TO CORRECTLY PACK BREATHING CIRCUITS. A PACK CARD DETAILING ALL COMPONENTS REQUIRED IS DISPLAYED AT THE PACKING STATION. EACH COMPLETED CIRCUIT PACK IS THEN WEIGHED TO ENSURE THAT EVERY COMPONENT CIRCUIT PACK IS ACCOUNTED FOR. (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT AN ADAPTOR WAS MISSING FROM AN RT329 INFANT BREATHING CIRCUIT KIT. THIS WAS FOUND BEFORE USE ON A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFANT CONTINUOUS FLOW BREATHING CIRCUIT BTT BTT FISHER & PAYKEL HEALTHCARE LTD RT329 101115

Patients

Seq Age Sex Outcome Treatment
1