FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2161092 · Received July 13, 2011

Report

Report Number
1423500-2011-09000
Event Type
Malfunction
Date Received
July 13, 2011
Date of Event
June 25, 2011
Report Date
June 25, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).THE SAMPLE WAS DISCARDED AND THE LOT NUMBER IS UNKNOWN. A BATCH REVIEW WILL NOT BE CONDUCTED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THEN A FOLLOW UP MDR WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 (AIR IN SET) OCCURRED DURING THE INITIAL DRAIN WAS NOT CONFIRMED DUE TO LACK OF SAMPLE. NOT ENOUGH DATA IS AVAILABLE WITHIN THE REPORT TO IDENTIFY ROOT CAUSE; THEREFORE, THE ROOT CAUSE OF THE COMPLAINT WAS NOT DETERMINED. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING THE HOME CHOICE (HC) SYSTEM ERROR (SE) 2240 (AIR IN THE SET), DURING THE INITIAL DRAIN. THE TECHNICAL SERVICES REPRESENTATIVE (TSR) EXPLAINED THE ALARM AND EXPLAINED THAT THE HOME PATIENT (HP) NEEDED TO END THERAPY AND START OVER WITH NEW SUPPLIES. THE TSR ASSISTED THE HP TO END THERAPY AND INSTRUCTED THE HP TO CALL HER NURSE OR A FAMILY MEMBER FOR HELP. BAXTER (B)(4) CONTACTED THE HP ON (B)(6) 2011. THE HP STATED SHE DID NOT NOTICE ANYTHING WITH THE SETUP AT THE TIME OF THE ALARM THAT SHE THOUGHT MAY HAVE CAUSED IT. THE HP STATED HER CG STARTED OVER WITH NEW SUPPLIES AND THE HP RESUMED THERAPY WITHOUT ANY PROBLEMS. THERE WAS PATIENT INVOLVEMENT. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 68 YR HOMECHOICE