FDA Adverse Event Malfunction Summary report: N

LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM VG

MDR report key: 2161086 · Received June 17, 2011

Report

Report Number
2024601-2011-00491
Event Type
Malfunction
Date Received
June 17, 2011
Date of Event
February 25, 2011
Report Date
May 24, 2011
Manufacturer
ALLERGAN
Product Code
LTI
PMA / PMN Number
P000008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). VISUAL EXAMINATION OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER II. ANALYSIS NOTED AN UNIDENTIFIED OPENING WITH LEAKAGE ON THE SHELL. THE SHELL/RING HAD A SHARP LINEAR OPENING WITH LEAKAGE, ETIOLOGY UNK. THE DAMAGED PORT SEPTUM HAD EVIDENCE OF CORING LIKELY TO HAVE BEEN CAUSED BY THE USE OF A CORING NEEDLE. THE ACCESS PORT HAD NON-PENETRATING NICKS WITH STRIATIONS DESCRIBED AS SURGICAL TOOL SCRATCH-LIKE. THE BAND TUBING WAS BROKEN WITH STRIATIONS CONSISTENT WITH A SURGICAL END CUT TO REMOVE THE DEVICE. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING."

Description of Event or Problem · 1

HEALTH PROFESSIONAL REPORTED ALLEGED LAP-BAND SYSTEM "EXPLANT DUE TO PORT LEAK." THE DEVICE WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM VG LTI ALLERGAN NA NI

Patients

Seq Age Sex Outcome Treatment
1 57 YR