LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM VG
Report
- Report Number
- 2024601-2011-00491
- Event Type
- Malfunction
- Date Received
- June 17, 2011
- Date of Event
- February 25, 2011
- Report Date
- May 24, 2011
- Manufacturer
- ALLERGAN
- Product Code
- LTI
- PMA / PMN Number
- P000008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). VISUAL EXAMINATION OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER II. ANALYSIS NOTED AN UNIDENTIFIED OPENING WITH LEAKAGE ON THE SHELL. THE SHELL/RING HAD A SHARP LINEAR OPENING WITH LEAKAGE, ETIOLOGY UNK. THE DAMAGED PORT SEPTUM HAD EVIDENCE OF CORING LIKELY TO HAVE BEEN CAUSED BY THE USE OF A CORING NEEDLE. THE ACCESS PORT HAD NON-PENETRATING NICKS WITH STRIATIONS DESCRIBED AS SURGICAL TOOL SCRATCH-LIKE. THE BAND TUBING WAS BROKEN WITH STRIATIONS CONSISTENT WITH A SURGICAL END CUT TO REMOVE THE DEVICE. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING."
HEALTH PROFESSIONAL REPORTED ALLEGED LAP-BAND SYSTEM "EXPLANT DUE TO PORT LEAK." THE DEVICE WAS REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM VG | LTI | ALLERGAN | NA | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR |