FDA Adverse Event
Malfunction
Summary report: N
LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (LARGE)
MDR report key: 2161085
·
Received June 17, 2011
Report
- Report Number
- 2024601-2011-00489
- Event Type
- Malfunction
- Date Received
- June 17, 2011
- Date of Event
- March 28, 2011
- Report Date
- May 23, 2011
- Manufacturer
- ALLERGAN
- Product Code
- LTI
- PMA / PMN Number
- P000008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). VISUAL EXAMINATION OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER II. ANALYSIS NOTED THE BAND TUBING WAS BROKEN WITH STRIATIONS CONSISTENT WITH A SURGICAL END CUT TO REMOVE THE DEVICE. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING."
Description of Event or Problem · 1
HEALTH PROFESSIONAL REPORTED, THE REPLACEMENT OF A PORT DUE TO AN ALLEGED LAP-BAND "PORT LEAK."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (LARGE) | LTI | ALLERGAN | NA | 1831263 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 36 YR |