FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2161036 · Received July 13, 2011

Report

Report Number
2531779-2011-04912
Event Type
Malfunction
Date Received
July 13, 2011
Report Date
June 15, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 06/29/2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. THE UP-ARROW AND BACKLIGHT KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS DURING TESTING. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER THE UP-ARROW AND BACKLIGHT KEY CONTACTS. (B)(6).

Description of Event or Problem · 1

A FAMILY MEMBER REPORTED THAT THE KEYPAD BUTTONS ARE SOMETIMES UNRESPONSIVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 6 YR