FDA Adverse Event
Malfunction
Summary report: N
ALLEGRETTO WAVE EYE-Q
MDR report key: 2160990
·
Received June 16, 2011
Report
- Report Number
- 3003288808-2011-00156
- Event Type
- Malfunction
- Date Received
- June 16, 2011
- Date of Event
- May 11, 2009
- Report Date
- May 11, 2009
- Manufacturer
- WAVELIGHT GMBH
- Product Code
- LZS
- PMA / PMN Number
- P030008/P0
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE SERVICE TECHNICIAN REPLACED THE WRP BOARD. AFTER REPLACEMENT, THE LASER WORKED W/O ANY PROBLEMS. ROOT CAUSE WAS A FAULTY WRP BOARD. (B)(4). THIS REPORT WAS MAILED TO FDA ON: (B)(4) 2011. (B)(4).
Description of Event or Problem · 1
CUSTOMER REPORTS THE DISTANCE DIODES DISAPPEAR OFTEN DURING SURGERY. NO PT INJURY WAS REPORTED AND NO FURTHER INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ALLEGRETTO WAVE EYE-Q | OPHTHALMIC EXCIMER LASER SYSTEM | LZS | WAVELIGHT GMBH | 8065990631 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |