VASOVIEW HEMOPRO EVH SYSTEM
Report
- Report Number
- 2242352-2011-00930
- Event Type
- Injury
- Date Received
- July 7, 2011
- Report Date
- June 13, 2011
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- GEI
- PMA / PMN Number
- K052274
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
INVESTIGATION: VISUAL INSPECTION REVEALED THAT THE C-RING HAD BEEN SPLIT IN HALF. THE OTHER HALF OF THE C-RING AND THE SCOPE WASH TUBING WERE RECEIVED SEPARATE. THE TUBING WAS INTACT. THE DEVICE WAS BLOODY. BASED UPON THE VISUAL OBSERVATIONS, THE REPORTED COMPLAINT "C-RING ARM BROKE OFF" WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. (B)(4).
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE VASOVIEW HEMOPRO C-RING ARM BROKE OFF INTO THE PATIENT'S LEG. THE PIECE WAS RETRIEVED THROUGH THE ORIGINAL INCISION. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS. THE PRODUCT WAS RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VASOVIEW HEMOPRO EVH SYSTEM | ENDOSCOPIC VESSEL HARVESTING SYSTEM | GEI | MAQUET CARDIOVASCULAR, LLC | VH-3000 | 25033953 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other |