Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). (ENDOLEAK) RESULTS/CONCLUSIONS: (TOO SHORT DISTAL LANDING ZONE).

Description of Event or Problem · 1

A TALENT THORACIC STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 6.5 CM ANEURYSM APPROXIMATELY TEN WEEKS AGO. VESSEL MORPHOLOGY WAS REPORTED AS A DISTAL DIAMETER OF APPROXIMATELY 28 MM; MODERATE CALCIFICATION IN THE DESCENDING THORACIC AORTA; NO ANGULATION. IT WAS REPORTED THAT THE DEVICE WAS DELIBERATELY LANDED SHORT OF THE LUMBAR ARTERIES (BETWEEN T10 AND T11). THE DISTAL SEAL LENGTH, AT IMPLANT, WAS LESS THAN 2 CM. THERE WAS NO ENDOLEAK AFTER IMPLANT. THE PATIENT DOES NOT APPEAR TO HAVE ANY MIGRATION; HOWEVER, A DISTAL TYPE I ENDOLEAK WAS SEEN ON RECENT FILMS. THE PATIENT IS ASYMPTOMATIC. THE PHYSICIAN PLANS TO PLACE AN EXTENSION AT A LATER DATE. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PATIENT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT CAPTIVIA STENT GRAFT MIH MEDTRONIC CARDIOVASCULAR IRELAND NA V00840146

Patients

Seq Age Sex Outcome Treatment
1 75 YR Required Intervention