TALENT CAPTIVIA STENT GRAFT
Report
- Report Number
- 2953200-2011-01245
- Event Type
- Injury
- Date Received
- July 7, 2011
- Date of Event
- June 7, 2011
- Report Date
- June 7, 2011
- Manufacturer
- MEDTRONIC CARDIOVASCULAR IRELAND
- Product Code
- MIH
- PMA / PMN Number
- P070007
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). (ENDOLEAK) RESULTS/CONCLUSIONS: (TOO SHORT DISTAL LANDING ZONE).
A TALENT THORACIC STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 6.5 CM ANEURYSM APPROXIMATELY TEN WEEKS AGO. VESSEL MORPHOLOGY WAS REPORTED AS A DISTAL DIAMETER OF APPROXIMATELY 28 MM; MODERATE CALCIFICATION IN THE DESCENDING THORACIC AORTA; NO ANGULATION. IT WAS REPORTED THAT THE DEVICE WAS DELIBERATELY LANDED SHORT OF THE LUMBAR ARTERIES (BETWEEN T10 AND T11). THE DISTAL SEAL LENGTH, AT IMPLANT, WAS LESS THAN 2 CM. THERE WAS NO ENDOLEAK AFTER IMPLANT. THE PATIENT DOES NOT APPEAR TO HAVE ANY MIGRATION; HOWEVER, A DISTAL TYPE I ENDOLEAK WAS SEEN ON RECENT FILMS. THE PATIENT IS ASYMPTOMATIC. THE PHYSICIAN PLANS TO PLACE AN EXTENSION AT A LATER DATE. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PATIENT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TALENT CAPTIVIA STENT GRAFT | MIH | MEDTRONIC CARDIOVASCULAR IRELAND | NA | V00840146 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Required Intervention |