NEXUS® BONESCALPEL® 10MM, BLUNT BLADE + IRRIGATION TUBING
Report
- Report Number
- 2435119-2025-00019
- Event Type
- Malfunction
- Date Received
- March 14, 2025
- Date of Event
- February 14, 2025
- Report Date
- March 14, 2025
- Manufacturer
- MISONIX, INC.
- Product Code
- LFL
- PMA / PMN Number
- K231117
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ON FEBRUARY 14, 2025, MISONIX® LLC., A BIOVENTUS® CO., RECEIVED A REPORT OF AN EVENT INVOLVING A NEXUS® BONESCALPEL® 10MM, BLUNT BLADE + IRRIGATION TUBING KIT (PART NUMBER: 110-31-1110, LOT NUMBER: 243269) THAT OCCURRED DURING A C2-T6 PSF PROCEDURE ON THE SAME DAY. SPECIFICALLY, THE REPORTER INDICATED THAT THE "SURGEON WAS SCORING BONE FOR CUTS TO PERFORM C3/C4 LAMI" AND THAT THE DISPOSABLE WAS "IN USE FOR 3 SECONDS BEFORE THE BLADE BROKE IN HALF." ALL FRAGMENTS OF THE BROKEN BLADE WERE RETRIEVED WITHOUT ADDITIONAL MEDICAL INTERVENTION. A SERIOUS INJURY TO THE PATIENT OR USER WAS NOT REPORTED. DELAY IN TREATMENT WAS NOT REPORTED. A REVIEW OF POST MARKET SURVEILLANCE DATA FOR THE NEXUS® SYSTEM DID NOT SHOW ANY SIGNIFICANT ADVERSE TRENDS FOR BREAKAGE OF ULTRASONIC PROBE TIPS. THE CURRENT FREQUENCY OF OCCURRENCE IS WITHIN THE FREQUENCY IN THE ORIGINAL RISK MANAGEMENT REPORT. THERE IS NO CHANGE TO THE RESIDUAL RISK OR RISK-BENEFIT RATIO OF THE DEVICE. THE RISK OF HARM IS MITIGATED BY THE NATURE IN WHICH TITANIUM ULTRASONIC BLADES BREAK DURING USE. TITANIUM DOES NOT SHATTER, SPLINTER, OR CREATE MULTIPLE FRAGMENTS THAT WOULD BE DIFFICULT TO LOCATE OR REMOVE FROM THE SURGICAL FIELD. IN MOST CASES, THE BROKEN PIECES CAN BE EASILY IDENTIFIED BY DIRECT VISUAL EXAMINATION AND REMOVED FROM THE SURGICAL FIELD. IN CASES WHERE BROKEN PIECES CANNOT BE IDENTIFIED VISUALLY, ENHANCED VISUALIZATION THROUGH THE ROUTINE USE OF LOUPE-FITTED EYEGLASSES OR MICROSCOPES CAN AID IN IDENTIFYING AND REMOVING BROKEN PIECES FROM THE SURGICAL FIELD. SURGICAL SUITES TYPICALLY HAVE ACCESS TO DIAGNOSTIC IMAGING EQUIPMENT, SUCH AS X-RAY OR FLUOROSCOPY, AND WOULD BE ABLE TO QUICKLY IDENTIFY AND LOCATE ANY FRAGMENTS NOT FOUND BY DIRECT OR ENHANCED VISUALIZATION. THE BROKEN PIECES THAT REMAIN IN THE OPERATIVE FIELD CAN THEREFORE BE FOUND AND EXTRACTED QUICKLY AND EASILY WITHOUT SIGNIFICANT DELAY IN THE SURGICAL PROCEDURE. THE INSTRUCTIONS FOR USE MANUAL (100-10-1000, REVISION L) FOR THE NEXUS® ULTRASONIC SURGICAL ASPIRATION SYSTEM CONTAINS THE FOLLOWING WARNINGS AND CAUTIONS TO PREVENT BLADE BREAKAGE. WARNING: THE NEXUS® ULTRASONIC SURGICAL ASPIRATOR SYSTEM IS INTENDED TO BE USED IN VARIOUS TYPES OF INVASIVE, SURGICAL PROCEDURES. THERE MAY BE INDIRECT DANGER TO THE PATIENT SHOULD THE DEVICE FAIL DURING THE PROCEDURE. IT IS RECOMMENDED THAT THE FACILITY FOLLOWS ITS BACK-UP EQUIPMENT PROTOCOLS. WARNING: ULTRASONIC PROBE TIPS CAN BREAK UNDER EXCESSIVE USE IN EXTREME CONDITIONS, E.G. WHEN CUTTING FOR EXTENDED DURATION IN TIGHT CAVITIES WITH LIMITED LATERAL MOTION. THE PROBE TIP COULD BREAK INTO TWO OR MORE FRAGMENTS WITH THE MAIN FRAGMENT REMAINING ATTACHED TO THE HANDPIECE. ALL FRAGMENTS MUST BE RETRIEVED IMMEDIATELY FROM THE SURGICAL SITE. THE FRAGMENTS SHOULD BE CHECKED TO ENSURE THAT NO FURTHER PIECES ARE MISSING. IT IS POSSIBLE THAT A FRAGMENT IS PROPELLED OUTSIDE OF THE SURGICAL CAVITY. DIAGNOSTIC IMAGING, SUCH AS X-RAY, MUST BE USED IF A FRAGMENT CANNOT BE FOUND TO CONFIRM THAT THE BROKEN PIECE IS OUTSIDE OF THE SURGICAL CAVITY. WARNING: BREAKAGE OF ULTRASONIC PROBE TIPS WILL RESULT IN SHARP EDGES THAT CAN BE HARMFUL TO SOFT TISSUE EVEN WITHOUT ACTIVATION OF ULTRASOUND. PROBE TIPS CAN BEND OR DEFORM BEFORE THEY ACTUALLY BREAK. PROBE TIPS SHOWING SIGNS OF DEFORMATION OR CRACKING SHOULD BE REPLACED IMMEDIATELY SINCE PROBE TIP BREAKAGE IS OTHERWISE IMMINENT. DO NOT BEND OR TWIST THE ULTRASONIC PROBE TIPS SINCE IT REDUCES THE STRUCTURAL INTEGRITY AND CAN RESULT IN PROBE TIP BREAKAGE DURING USE. DISPOSE OF DEFORMED OR BROKEN PROBE TIPS IMMEDIATELY IN A SHARPS CONTAINER IN ACCORDANCE WITH YOUR FACILITY BIOLOGICAL HAZARDOUS WASTE PROCEDURE. WARNING: DURING SYSTEM CHECK, MAKE SURE THE PROBE TIP OF THE HANDPIECE IS FREE FROM CONTACT WITH ANY OBJECT. ALLOWING CONTACT WITH THE PROBE TIP MAY RESULT IN DAMAGE AND/OR PERSONAL INJURY. CAUTION: THE SYSTEM CHECK SHOULD ALWAYS BE DONE IN ADVANCE OF PREPARING PATIENT FOR SURGERY TO MINIMIZE RISK TO PATIENT IN CASE OF SYSTEM MALFUNCTION. REFER TO SECTION 7.11 - SYSTEM CHECK. CAUTION: LOOSE PROBE TIP/TISSUE CONTACT UPON AN INITIAL BONE INCISION CAN CAUSE A THIN PROBE TIP TO RESONATE NOT ONLY LONGITUDINALLY BUT ALSO TRANSVERSELY. THIS CAN CAUSE A THIN PROBE TIP TO BREAK. IT IS NECESSARY TO ENGAGE BONE ACTIVELY AND WITH A MINIMAL PROBE TIP PRESSURE GREATER THAN ZERO IN ORDER TO PREVENT THE SHATTERING. CAUTION: CONTACT OF THE ULTRASONIC PROBE TIP OR THE EXPOSED EXTENSION WITH METAL, SURGICAL INSTRUMENTS OR OTHER OBJECTS DURING ULTRASOUND USE MUST BE AVOIDED. SUCH CONTACT CAN DAMAGE THE ULTRASONIC COMPONENTS VERY EASILY AND MAY RESULT IN COMPROMISED PERFORMANCE, INCLUDING FAILURE. DISCARD ANY EXTENSIONS OR PROBE TIPS THAT SHOW SIGNS OF DAMAGES LIKE GOUGES, NICKS OR FRACTURES. EXTERNAL ASPIRATION MAY BE USED BUT IT IS RECOMMENDED THAT A PLASTIC SUCTION PROBE TIP SHOULD BE USED WHEN IN PROXIMITY WITH THE PROBE TIP. CAUTION: THE NEXUS® SYSTEM SHOULD BE FULLY TESTED AND INSPECTED PRIOR TO EACH PROCEDURE. THE CONSOLE, FOOTSWITCH, HANDPIECES, ALL CABLES AND ACCESSORIES SHOULD BE EXAMINED FOR PROPER APPEARANCE AND CONDITION. THE SUBJECT DEVICE HAS NOT YET BEEN RETURNED TO MISONIX FOR EVALUATION, THEREFORE A THOROUGH INVESTIGATION COULD NOT BE CONDUCTED. UPON RETURN A FULL ENGINEERING EVALUATION WILL BE COMPLETED TO ESTABLISH POSSIBLE ROOT CAUSE OF THE NON-CONFORMANCE.
ON FEBRUARY 14, 2025, MISONIX® LLC., A BIOVENTUS® CO., RECEIVED A REPORT OF AN EVENT INVOLVING A NEXUS® BONESCALPEL® 10MM, BLUNT BLADE + IRRIGATION TUBING KIT (PART NUMBER: 110-31-1110, LOT NUMBER: 243269) THAT OCCURRED DURING A C2-T6 PSF PROCEDURE ON THE SAME DAY. SPECIFICALLY, THE REPORTER INDICATED THAT THE "SURGEON WAS SCORING BONE FOR CUTS TO PERFORM C3/C4 LAMI" AND THAT THE DISPOSABLE WAS "IN USE FOR 3 SECONDS BEFORE THE BLADE BROKE IN HALF." ALL FRAGMENTS OF THE BROKEN BLADE WERE RETRIEVED WITHOUT ADDITIONAL MEDICAL INTERVENTION. A SERIOUS INJURY TO THE PATIENT OR USER WAS NOT REPORTED. DELAY IN TREATMENT WAS NOT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1229014 | NEXUS® BONESCALPEL® 10MM, BLUNT BLADE + IRRIGATION TUBING | 10MM BLUNT BLADE AND IRRIGATION TUBING KIT | LFL | MISONIX, INC. | 110-31-1110 | 243269 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Other |