LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2011-00120
- Event Type
- Malfunction
- Date Received
- July 1, 2011
- Date of Event
- May 23, 2011
- Report Date
- June 29, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DAMAGED CONNECTOR) HAS BEEN CONFIRMED. UPON EVALUATION THE TRUNK CABLE CONNECTOR WAS DAMAGED. THE CAUSE FOR THE DAMAGED CONNECTOR CANNOT BE POSITIVELY DETERMINED BUT WAS LIKELY THE RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT CONNECTOR. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.
A (B)(6) MALE PATIENT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT HIS MONITOR IS DISPLAYING A MESSAGE STATING HIS ELECTRODE BELT IS NOT CONNECTED. HE COULD NOT TROUBLESHOOT THE ISSUE. THE PATIENT ADMITTED THAT HE BENT OVER AT THE WAIST AND THE ELECTRODE BELT WAS PULLED. THE PATIENT WAS PROVIDED WITH A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR |