Report

Report Number
1723170-2025-01302
Event Type
Malfunction
Date Received
March 14, 2025
Date of Event
March 6, 2025
Report Date
April 22, 2025
Manufacturer
MEDTRONIC NAVIGATION, INC
Product Code
HAW
PMA / PMN Number
K162309
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PO
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H2) ADDITIONAL INFORMATION WAS ADDED TO B5. H3) THE MANUFACTURER REPRESENTATIVE WENT TO THE SITE TO TEST THE NAVIGATION SYSTEM. THERE WAS A LOCALIZER FAULT ON AND AN AMBER LIGHT ON THE CAMERA. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H2: CONCOMITANT PRODUCT: 9735821. LOT #: P9-02327. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

D10: CONCOMITANT PRODUCT: PRODUCT ID: 9735821, MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

B3: DATE IS APPROXIMATE. MONTH AND YEAR ARE CONFIRMED VALID. H3) A MANUFACTURER REPRESENTATIVE WENT TO THE SITE TO TEST THE NAVIGATION SYSTEM. THE CAMERA WAS NOT WORKING PROPERLY AND NEEDED TO BE REPLACED. CODES: B01, C08, D02. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE REPORTED EVENT OCCURRED ON MARCH 06, 2025. THE ISSUE OCCURRED PRE-OPERATIVELY.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED IN AN UNKNOWN CONTEXT. IT WAS REPORTED THAT "LOCALIZER FAULTED" ERROR MESSAGE WAS IN OPTICAL PROCEDURES. THE NAVIGATION CAMERA WAS NOT WORKING PROPERLY. PATIENT PRESENCE UNKNOWN. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1616091 STEALTHSTATION¿ S8 SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT HAW MEDTRONIC NAVIGATION, INC 9735665

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown SEE H11.