FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 2160482
·
Received July 1, 2011
Report
- Report Number
- 3008642652-2011-00126
- Event Type
- Malfunction
- Date Received
- July 1, 2011
- Date of Event
- May 21, 2011
- Report Date
- June 30, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (SERVICE CODE 204) HAS BEEN CONFIRMED. UPON EVALUATION, THE TRUNK CABLE CONNECTOR WAS DAMAGED. THE CAUSE FOR THE DAMAGED CONNECTOR CANNOT BE POSITIVELY DETERMINED BUT WAS LIKELY THE RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT CONNECTOR. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
THE REVIEW OF A (B)(6) FEMALE PATIENT REVEALED SEVERAL CAN COMM FAULTS, BELT COMMUNICATION ERRORS, AND ABNORMAL SHUTDOWNS. ZOLL CUSTOMER SUPPORT CONTACTED THE PATIENT TO DISCUSS THE DOWNLOAD. AS A RESULT OF THE CALL, THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR |