FDA Adverse Event
Malfunction
Summary report: N
ESPRIT VENTILATOR
MDR report key: 2160451
·
Received July 13, 2011
Report
- Report Number
- 2031642-2011-00225
- Event Type
- Malfunction
- Date Received
- July 13, 2011
- Date of Event
- June 22, 2011
- Report Date
- June 22, 2011
- Manufacturer
- RESPIRONICS CALIFORNIA, INC.
- Product Code
- CBK
- PMA / PMN Number
- K981072
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Description of Event or Problem · 1
FACTORY FAILURE ANALYSIS OF THE MAIN PCB BOARD REVEALED PHYSICAL DAMAGE AND A BROKEN SOLDER CONNECTION. THE VENTILATOR'S REMOTE ALARM PORT PROVIDES A SIGNAL DURING VENTILATOR ALARM CONDITIONS TO BE SOUNDED AT REMOTE LOCATIONS AWAY FROM THE VENTILATOR. A BROKEN SOLDER CONNECTION OF THE REMOTE ALARM CONNECTOR MAY RESULT IN NO ALARM SOUNDING TO THE CUSTOMER AT A REMOTE STATION IF THERE IS A VENTILATOR ALARM CONDITION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ESPRIT VENTILATOR | VENTILATOR, CONTINUOUS | CBK | RESPIRONICS CALIFORNIA, INC. | V1000 | NA |