FDA Adverse Event Malfunction Summary report: N

ESPRIT VENTILATOR

MDR report key: 2160451 · Received July 13, 2011

Report

Report Number
2031642-2011-00225
Event Type
Malfunction
Date Received
July 13, 2011
Date of Event
June 22, 2011
Report Date
June 22, 2011
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
CBK
PMA / PMN Number
K981072
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

FACTORY FAILURE ANALYSIS OF THE MAIN PCB BOARD REVEALED PHYSICAL DAMAGE AND A BROKEN SOLDER CONNECTION. THE VENTILATOR'S REMOTE ALARM PORT PROVIDES A SIGNAL DURING VENTILATOR ALARM CONDITIONS TO BE SOUNDED AT REMOTE LOCATIONS AWAY FROM THE VENTILATOR. A BROKEN SOLDER CONNECTION OF THE REMOTE ALARM CONNECTOR MAY RESULT IN NO ALARM SOUNDING TO THE CUSTOMER AT A REMOTE STATION IF THERE IS A VENTILATOR ALARM CONDITION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ESPRIT VENTILATOR VENTILATOR, CONTINUOUS CBK RESPIRONICS CALIFORNIA, INC. V1000 NA