FDA Adverse Event
Malfunction
Summary report: N
TERUMO TCM II COOLING AND HEATING SYSTEM
MDR report key: 2160419
·
Received June 20, 2011
Report
- Report Number
- 1828100-2011-01715
- Event Type
- Malfunction
- Date Received
- June 20, 2011
- Date of Event
- May 20, 2011
- Report Date
- June 20, 2011
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP.
- Product Code
- DWC
- PMA / PMN Number
- K883603
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
UPON INSTALLATION OF THE DEVICE, THE USER REPORTED THAT THE DEVICE BOOTED UP, BUT THERE WAS NO REACTION WHEN THE COOL DOWN AND CARDIOPLEGIA WERE ACTIVATED. INVESTIGATION AT USER FACILITY REVEALED AN IMPROPER CONNECTION OF CONNECTIONS 13 AND 14. THE DEVICE WAS REPAIRED AT THE USER FACILITY AND WILL NOT BE RETURNED. SINCE THE EVENT OCCURRED UPON INSTALLATION, THERE WAS NO PATIENT INVOLVEMENT DURING THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TERUMO TCM II COOLING AND HEATING SYSTEM | HEATING COOLING SYSTEM | DWC | TERUMO CARDIOVASCULAR SYSTEMS CORP. | 4415 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |