FDA Adverse Event Malfunction Summary report: N

TERUMO TCM II COOLING AND HEATING SYSTEM

MDR report key: 2160419 · Received June 20, 2011

Report

Report Number
1828100-2011-01715
Event Type
Malfunction
Date Received
June 20, 2011
Date of Event
May 20, 2011
Report Date
June 20, 2011
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DWC
PMA / PMN Number
K883603
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

UPON INSTALLATION OF THE DEVICE, THE USER REPORTED THAT THE DEVICE BOOTED UP, BUT THERE WAS NO REACTION WHEN THE COOL DOWN AND CARDIOPLEGIA WERE ACTIVATED. INVESTIGATION AT USER FACILITY REVEALED AN IMPROPER CONNECTION OF CONNECTIONS 13 AND 14. THE DEVICE WAS REPAIRED AT THE USER FACILITY AND WILL NOT BE RETURNED. SINCE THE EVENT OCCURRED UPON INSTALLATION, THERE WAS NO PATIENT INVOLVEMENT DURING THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TERUMO TCM II COOLING AND HEATING SYSTEM HEATING COOLING SYSTEM DWC TERUMO CARDIOVASCULAR SYSTEMS CORP. 4415

Patients

Seq Age Sex Outcome Treatment
1