FDA Adverse Event Malfunction Summary report: N

6800

MDR report key: 2160409 · Received June 20, 2011

Report

Report Number
1720753-2011-08287
Event Type
Malfunction
Date Received
June 20, 2011
Date of Event
May 27, 2011
Report Date
June 20, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE SERVICE REPRESENTATIVE RESEATED THE MONOBLOCK, TRANSITION CABLE, AND THE TRANSITION AND CONTROLLER PRINTED CIRCUIT BOARD. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD DISPLAY A SYSTEM ERROR MESSAGE AND REQUIRE A REBOOT. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 6800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 6800

Patients

Seq Age Sex Outcome Treatment
1