DORNIER HOLMIUM LASER FIBER
Report
- Report Number
- 1037955-2011-00008
- Event Type
- Malfunction
- Date Received
- May 16, 2011
- Report Date
- May 16, 2011
- Manufacturer
- DORNIER MEDTECH AMERICA, INC.
- Product Code
- GEX
- PMA / PMN Number
- 022544
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
EVALUATION OF THIS COMPLAINT PRODUCT INDICATES THAT THE FIBER BREAK LIKELY OCCURRED DUE TO A SEVERE BEND BEYOND THE FIBER MINIMUM BEND RADIUS OF 7MM WHILE UNDER LASER FIRE. THIS BEND LIKELY OCCURRED FROM THE APPLICATION OF EXTERNAL MECHANICAL FORCE TO THE FIBER DIRECTLY BEHIND THE REAR STAINLESS STEEL SMA PIN CAUSING THE FIBER TO BE BENT BEYOND THE MINIMUM BEND RADIUS WHILE THE FIBER WAS UNDER FIRE BY A LASER. IT WAS FURTHER HYPOTHESIZED THAT THE OPERATING ROOM TECH WAS BURNED ON HER ABDOMEN BY THE BREAKING POINT OF THE FIBER BECAUSE HER ABDOMEN MAY HAVE ACTED AS THE EXTERNAL MECHANICAL FORCE WHICH WAS APPLIED TO THE FIBER WHILE UNDER LASER FIRE CAUSING THE BREAK (THE OPERATING ROOM TECH LEANED OVER THE LASER WITH HER ABDOMEN HITTING THE FIBER WHILE UNDER FIRE BENDING THE FIBER CAUSING THE BREAK.) THIS SPECULATION IS SUPPORTED BY OBJECTIVE EVIDENCE THAT THE SMA CONNECTION POINT ON A LASER IS POSITIONED AT ABDOMEN HEIGHT ON MOST HOLMIUM LASERS AND THAT DIRECT TISSUE BURNS FROM 270 UM DORNIER FIBERS AT THE MAXIMUM FIBER CAPABILITY OF 20 WATTS CAN ONLY OCCUR WITHIN A RANGE OF ONE TO TWO INCHES OF THE LASER ENERGY EXITING THE TIP OF A FIBER, OR IN THIS CASE THE BREAKING POINT OF THE FIBER. BURN TESTING FOR THIS DISTANCE DETERMINATION WAS PERFORMED WITH CALIBRATED TEST AND MEASUREMENT EQUIPMENT. THIS FIBER BEND (BREAK) COULD ONLY HAVE OCCURRED DURING FIBER USE BY THE END USER AS SUCH AN EVIDENT BEND (BREAK) WOULD HAVE FAILED POWER TESTING DURING FIBER MANUFACTURING IN ADDITION TO BEING CLEARLY NOTED BY FIBER PRODUCTION PERSONNEL AS AN ABNORMAL DEFECT. THIS FIBER BREAK APPEARS TO BE FROM A USER HANDLING ISSUE.
FIBER BROKE AT SMA CONNECTOR TO LASER. OPERATING ROOM TECH'S GOWN AND ABDOMEN WAS BURNT. NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DORNIER HOLMIUM LASER FIBER | LASER FIBER | GEX | DORNIER MEDTECH AMERICA, INC. | DUR270D | F0411A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |