THERAPY COOL FLEX, 1304-CF-7-0.5(5)2-L-TE4BE1EB
Report
- Report Number
- 2030404-2011-00116
- Event Type
- Malfunction
- Date Received
- April 29, 2011
- Date of Event
- April 5, 2011
- Report Date
- April 5, 2011
- Manufacturer
- ST. JUDE MEDICAL, IRVINE
- Product Code
- OAD
- PMA / PMN Number
- P060019
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THE CATHETER WAS NOT RETURNED FOR EVALUATION. REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE DEVICE MET MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. INFORMATION IN THE COMPLAINT DESCRIPTION IS CONSISTENT WITH A DETACHED LURE EXTENSION TUBE. A QUALITY PROJECT HAS BEEN INSTITUTED THAT CONSISTS OF IMPROVEMENTS TO THE TUBING QUALITY AND MORE IMPROVEMENTS TO THE INTERNAL INSPECTION PROCESS. SJM PARTNERED WITH THE VENDOR OF THE TUBING AND DETERMINED THAT THE MATERIAL WAS NOT PROCESSED WITH THE NECESSARY CONDITIONS TO ENSURE OPTIMUM TUBING QUALITY; THE PROCESS HAS BEEN IMPROVED BY THE VENDOR. IN ORDER BETTER PREVENT POTENTIALLY DEFECTIVE TUBING FROM BEING UTILIZED DURING THE MANUFACTURING PROCESS, SJM HAS ADDITIONALLY INSTITUTED A NEW TESTING FIXTURE THAT IMPROVES OUR ABILITY TO DETECT THE NONCONFORMING TUBING.
SAME CASE AS MFR. REPORT #2030404-2011-00015. IT WAS REPORTED THAT DURING A PULMONARY VEIN ABLATION PROCEDURE, THE IRRIGATION PORT DISCONNECTED FROM THE HANDLE OF THE CATHETER. WHILE USING A THERAPY COOL FLEX CATHETER, THE PHYSICIAN NOTED THAT THE IRRIGATION PORT ON THE PROXIMAL END HAD DISCONNECTED AND WAS NOT IRRIGATING THE CATHETER TIP. THE CATHETER WAS EXCHANGED FOR A COOL PATH DUO THAT DEVELOPED AN 'S' CURVE WHEN FLEXED AFTER ABOUT 45 MINUTES OF USE. THE CATHETER WAS EXCHANGED FOR A SECOND COOL PATH DUO AND WHILE ABLATING AT THE CAVOTRICUSPID ISTHMUS (CTI) FOR APPROXIMATELY 10 MINUTES A "LOUD STEAM POP" WAS HEARD. THE PATIENT REMAINED HEMODYNAMICALLY STABLE SO THE PHYSICIAN CONTINUED ABLATING FOR ANOTHER 5 MINUTES. THE PHYSICIAN STOPPED ABLATING BEFORE ACHIEVING CTI BLOCK AND ENDED THE PROCEDURE WITH NO CONSEQUENCES TO THE PATIENT. THE PHYSICIAN CONSIDERED THIS A SUCCESSFUL ABLATION PROCEDURE DESPITE NOT ACHIEVING BIDIRECTIONAL BLOCK ACROSS THE CTI.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | THERAPY COOL FLEX, 1304-CF-7-0.5(5)2-L-TE4BE1EB | CARDIAC ABLATION CATHETER PERCUTANEOUS | OAD | ST. JUDE MEDICAL, IRVINE | 88016 | 3223034 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | SJM AGILIS INTRODUCER:| COOL PATH DUO: MODEL 83563, LOT 3189219| COOL PATH DUO: MODEL 88016, LOT 3223034 |