COULTER LH 750 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2011-00543
- Event Type
- Malfunction
- Date Received
- June 2, 2011
- Date of Event
- August 25, 2008
- Report Date
- September 8, 2008
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- GKZ
- PMA / PMN Number
- K061574
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HK
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
FIELD SERVICE INFORMATION WAS NOT AVAILABLE. RAW DATA ANALYSIS WAS CONDUCTED AND DETERMINED THAT THE WBC HISTOGRAMS SHOWED SEVERE INTERFERENCE PATTERNS. ROOT CAUSE WAS DETERMINED TO BE UNLYSED RED BLOOD CELLS IN THE WBC HISTOGRAM CAUSING SEVERE INTERFERENCE. PRODUCT LABELING STATES THAT INSTRUMENT GENERATED MESSAGES ALERT THE OPERATOR TO FURTHER REVIEW THE SAMPLE. LYSE RESISTANT RED CELLS ARE LISTED AS KNOWN INTERFERING SUBSTANCES FOR THE WBC PARAMETER. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW CONDUCTED BETWEEN (B)(4), 2008 AND (B)(4), 2010 OF COMPLAINTS FOR ADDITIONAL REPORTABLE EVENTS. THIS MDR REPRESENTS EVENT 2 OF 3 REPORTED BY THIS CUSTOMER. THIS MDR IS RELATED TO THE FOLLOWING MDRS THAT HAVE BEEN REPORTED: MDR 1061932-2011-00541, 1061932-2011-00544.
ON (B)(6) 2008, CUSTOMER REPORTED ELEVATED AND INCONSISTENT UNCORRECTED WBC (WHITE BLOOD COUNT) FOR A SPECIFIC SAMPLE WHEN USING THE COULTER LH 750 HEMATOLOGY ANALYZER. THE TEST RESULTS WERE GENERATED ON (B)(6) 2008 AND WERE FLAGGED WITH INSTRUMENT GENERATED MESSAGES FOR THE WBC PARAMETER. THE WBC TEST RESULTS WERE COMPARABLE TO A MANUAL COUNT OF 9.0 X 10 TO THE 3RD POWER CELLS/UL. THE MANUAL COUNT WAS REPORTED AS CORRECT. ERRONEOUS RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. NO REPORTS OF DEATH OR SERIOUS INJURY, AND NO AFFECT TO PATIENT TREATMENT AS A RESULT OF THIS EVENT. THIS EVENT REPRESENTS EVENT 2 OF 3 EVENTS REPORTED BY THIS CUSTOMER FOR THE SAME PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COULTER LH 750 HEMATOLOGY ANALYZER | GKZ | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |