FDA Adverse Event Injury Summary report: N

HEARTMATE II LVAS

MDR report key: 2160247 · Received July 13, 2011

Report

Report Number
2160247
Event Type
Injury
Date Received
July 13, 2011
Date of Event
April 27, 2011
Report Date
September 27, 2011
Manufacturer
THORATEC CORPORATION
Product Code
DSQ
Adverse Event
Yes
Product Problem
Yes
Report Source
User Facility report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

MAJOR PUMP UNIT(S) INVOLVED: EXTERNAL CONTROL SYSTEM FAILUREADDITIONAL TEXT: VISUAL AND AUDIBLE ALARMS DIDN'T MATCHSPECIFIC COMPONENT(S) INVOLVED: EXTERNAL CONTROLLER MALFUNCTIONADDITIONAL TEXT: VISUAL AND AUDIBLE ALARMS DIDN'T MATCHOTHER COMPONENT:CAUSATIVE OR CONTRIBUTING FACTOR: NO SPECIFIC CONTRIBUTING CAUSE IDENTIFIEDOTHER CAUSE:INTERVENTION(S): REPLACEMENT OF INTERNAL CONTROLLEROTHER INTERVENTION :IMPLANT DEVICE TYPE: LVADMALFUNCTION DEVICE TYPE:

Description of Event or Problem · 1

MAJOR PUMP UNIT(S) INVOLVED: EXTERNAL CONTROL SYSTEM FAILURE. SPECIFIC COMPONENT(S) INVOLVED: EXTERNAL CONTROLLER MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTMATE II LVAS LVAD DSQ THORATEC CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 22.4 YR