VASOVIEW HEMOPRO EVH SYSTEM
MDR report key: 2160182
·
Received May 13, 2011
Report
- Report Number
- 2242352-2011-00443
- Event Type
- Malfunction
- Date Received
- May 13, 2011
- Date of Event
- April 19, 2011
- Report Date
- April 20, 2011
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- GEI
- PMA / PMN Number
- K052274
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATION: VISUAL INSPECTION REVEALED THAT TIP OF THE COLD JAW SILICONE BOOT HAD DETACHED AND WAS PEELING. THE JAW HAD SOME SIGNS OF USAGE. THERE WAS SOME EVIDENCE OF BLOOD. BASED UPON THE VISUAL OBSERVATIONS, THE REPORTED COMPLAINT FOR "JAW BOOT DETACHED" WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. (B)(4).
Description of Event or Problem · 1
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE VASOVIEW HEMOPRO JAW BOOT DETACHED. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THE PRODUCT WAS RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VASOVIEW HEMOPRO EVH SYSTEM | ENDOSCOPIC VESSEL HARVESTING SYSTEM | GEI | MAQUET CARDIOVASCULAR, LLC | VH-3000 | 25030397 |