VASOVIEW HEMOPRO EVH SYSTEM

MDR report key: 2160182 · Received May 13, 2011

Report

Report Number
2242352-2011-00443
Event Type
Malfunction
Date Received
May 13, 2011
Date of Event
April 19, 2011
Report Date
April 20, 2011
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
GEI
PMA / PMN Number
K052274
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION: VISUAL INSPECTION REVEALED THAT TIP OF THE COLD JAW SILICONE BOOT HAD DETACHED AND WAS PEELING. THE JAW HAD SOME SIGNS OF USAGE. THERE WAS SOME EVIDENCE OF BLOOD. BASED UPON THE VISUAL OBSERVATIONS, THE REPORTED COMPLAINT FOR "JAW BOOT DETACHED" WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE VASOVIEW HEMOPRO JAW BOOT DETACHED. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THE PRODUCT WAS RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VASOVIEW HEMOPRO EVH SYSTEM ENDOSCOPIC VESSEL HARVESTING SYSTEM GEI MAQUET CARDIOVASCULAR, LLC VH-3000 25030397

Patients

Seq Age Sex Outcome Treatment
1 NA