FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2160167 · Received July 13, 2011

Report

Report Number
6000001-2011-11672
Event Type
Malfunction
Date Received
July 13, 2011
Date of Event
June 16, 2011
Report Date
June 16, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A BATTERY ISSUE WAS CONFIRMED DURING PRODUCT EVALUATION. THE ROOT CAUSE OF THIS CONDITION WAS ASSIGNED TO DEPLETED MAIN BATTERIES RESULTING FROM USER ERROR. THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED TO CORRECT THIS CONDITION. ADDITIONAL INFORMATION: THIS INVOLVED A COLLEAGUE P1.5 INFUSION PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.13.92. A LABELING REVIEW HAS BEEN PERFORMED, FINDING THAT CURRENT LABELING PROVIDES AMPLE INSTRUCTIONS RELATED TO PREVENTION OF THE SUSPECTED USER ERROR.

Additional Manufacturer Narrative · 1

THE DEVICE WAS EVALUATED ON-SITE AT THE CUSTOMER FACILITY. BAXTER'S INVESTIGATION IS STILL IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE. (B)(4).

Description of Event or Problem · 1

THIS IS A REPORT FROM BAXTER (B)(4) OF A COLLEAGUE INFUSION PUMP WITH A BATTERY ISSUE. IT IS UNKNOWN WHEN THE REPORTED CONDITION OCCURRED. IT IS UNKNOWN IF THERE IS PATIENT INVOLVEMENT. NO PATIENT INJURY OR MEDICAL INTERVENTION OCCURRED. THE SOFTWARE VERSION IS CURRENTLY NOT KNOWN. NO ADDITIONAL INFORMATION IS AVAILABLE.