FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2160064 · Received July 13, 2011

Report

Report Number
2531779-2011-04891
Event Type
Malfunction
Date Received
July 13, 2011
Report Date
June 15, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED BACK TO ANIMAS WITH THE FOLLOWING FINDINGS. THE TESTING CONFIRMED INTERMITTENT RESPONSES TO BUTTON PRESSES ON THE UP ARROW BUTTON AND THE CONTRAST BUTTON. PRODUCT ANALYSIS REMOVED THE KEYPAD COVER TO CHECK CONDITION OF THE BUTTON CONTACTS AND NOTICED THAT THERE WAS CONTAMINATION UNDER THE CONTACTS OF ALL BUTTONS. PRODUCT ANALYSIS ALSO NOTED THAT THE BUTTON CONTACT ON THE UP ARROW BUTTON WAS INVERTED.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 08/08/2011. DEVICE HISTORY RECORD REVIEW WAS CONDUCTED ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PATIENT CLAIMED THAT THE UP BUTTONS REQUIRED SEVERAL PRESSES TO ACTIVATE THE DESIRED PUMP FUNCTIONS. THE BACK LIGHT WAS ALSO HARD TO RESPOND. THE KEYPAD IS REPORTEDLY INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATE WITH THIS COMPLAINT. THE PUMP WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 53 YR