ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2011-04891
- Event Type
- Malfunction
- Date Received
- July 13, 2011
- Report Date
- June 15, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP WAS RETURNED BACK TO ANIMAS WITH THE FOLLOWING FINDINGS. THE TESTING CONFIRMED INTERMITTENT RESPONSES TO BUTTON PRESSES ON THE UP ARROW BUTTON AND THE CONTRAST BUTTON. PRODUCT ANALYSIS REMOVED THE KEYPAD COVER TO CHECK CONDITION OF THE BUTTON CONTACTS AND NOTICED THAT THERE WAS CONTAMINATION UNDER THE CONTACTS OF ALL BUTTONS. PRODUCT ANALYSIS ALSO NOTED THAT THE BUTTON CONTACT ON THE UP ARROW BUTTON WAS INVERTED.
FOLLOW-UP # 1 08/08/2011. DEVICE HISTORY RECORD REVIEW WAS CONDUCTED ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PATIENT CLAIMED THAT THE UP BUTTONS REQUIRED SEVERAL PRESSES TO ACTIVATE THE DESIRED PUMP FUNCTIONS. THE BACK LIGHT WAS ALSO HARD TO RESPOND. THE KEYPAD IS REPORTEDLY INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATE WITH THIS COMPLAINT. THE PUMP WAS REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR |