FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2160023 · Received July 13, 2011

Report

Report Number
2531779-2011-04888
Event Type
Malfunction
Date Received
July 13, 2011
Report Date
June 14, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD BUTTONS WERE FOUND TO BE RESPONDING APPROPRIATELY TO BUTTON PRESSES. THE BOLUS BUTTON WAS FOUND TO BE DETACHED. THE BOLUS BUTTON WAS FOUND TO BE LEAKING AND THERE WAS INTERNAL MOISTURE OBSERVED NEAR THE BOLUS BUTTON.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE BUTTONS ARE NOT RESPONDING AND THE AUDIO BOLUS BUTTON IS PEELING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 1 YR