FDA Adverse Event Malfunction Summary report: N

ALINITY I FREE T4 REAGENT KIT

MDR report key: 21599763 · Received March 13, 2025

Report

Report Number
3005094123-2025-00112
Event Type
Malfunction
Date Received
March 13, 2025
Date of Event
March 3, 2025
Report Date
May 30, 2025
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
CEC
UDI-DI
00380740173715
PMA / PMN Number
K173122
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE COMPLAINT INVESTIGATION INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, LABELING REVIEW, DEVICE HISTORY RECORD REVIEW, AND IN HOUSE TESTING. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE WITHOUT INDICATION FOR ANY ADDITIONAL ISSUE. A REVIEW OF TRACKING AND TRENDING DID IDENTIFY AN INCREASE IN COMPLAINTS FOR THE ALINITY I FREE T4 REAGENT AND FOR THE COMPLAINT LOT. HOWEVER, IN-HOUSE TESTING WAS COMPLETED WHICH CONCLUDED THE COMPLAINT LOT MET ACCEPTANCE CRITERIA, DEMONSTRATING THAT THE LOT IS PERFORMING AS EXPECTED. DEVICE HISTORY REVIEW DID NOT IDENTIFY ANY ISSUES WITH THE COMPLAINT LOT. LABELING WAS REVIEWED AND WAS FOUND TO ADDRESS THE CUSTOMER REPORTED ISSUE. BASED ON THE AVAILABLE INFORMATION, NO SYSTEMIC ISSUE OR DEFICIENCY WAS IDENTIFIED FOR THE ALINITY I FREE T4 REAGENT LOT 65500UD03 WAS IDENTIFIED.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 07P70-77 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 7P70, WITH 510K/PMA/BLA NUMBER K173122. AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem · 0

THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY I FREE T4 AND PROVIDED THE FOLLOWING DATA: INITIAL TEST RESULT WAS 31.14 PMOL/L, RE-TEST RESULT WAS 17.08 PMOL/L (CUSTOMER REFERENCE RANGE: 9.01-19.05) PER THE CUSTOMER DISCREPANT RESULTS WERE NOT REPORTED OUT TO THE MEDICAL PROVIDER. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

Description of Event or Problem · 0

THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY I FREE T4 AND PROVIDED THE FOLLOWING DATA: INITIAL TEST RESULT WAS 31.14 PMOL/L, RE-TEST RESULT WAS 17.08 PMOL/L (CUSTOMER REFERENCE RANGE: 9.01-19.05). PER THE CUSTOMER DISCREPANT RESULTS WERE NOT REPORTED OUT TO THE MEDICAL PROVIDER. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
804717 ALINITY I FREE T4 REAGENT KIT RADIOIMMUNOASSAY, FREE THYROXINE CEC ABBOTT IRELAND DIAGNOSTICS DIVISION 65500UD03 00380740173715

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown ALNTY I PROCESSING MODU, 03R65-01, (B)(6).| ALNTY I PROCESSING MODU, 03R65-01, (B)(6).