ARCHITECT FREE T4 REAGENT KIT
Report
- Report Number
- 3005094123-2025-00111
- Event Type
- Malfunction
- Date Received
- March 13, 2025
- Date of Event
- March 2, 2025
- Report Date
- May 30, 2025
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- CEC
- UDI-DI
- 00380740173678
- PMA / PMN Number
- K173122
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE COMPLAINT INVESTIGATION INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, LABELING REVIEW, DEVICE HISTORY RECORD REVIEW, AND IN HOUSE TESTING. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE WITHOUT INDICATION FOR ANY ADDITIONAL ISSUE. A REVIEW OF TRACKING AND TRENDING DID IDENTIFY AN INCREASE IN COMPLAINTS FOR THE ARCHITECT FREE T4 REAGENT AND FOR THE COMPLAINT LOT 68900UD02. HOWEVER, IN-HOUSE TESTING WAS COMPLETED WHICH CONCLUDED THE COMPLAINT LOT MET ACCEPTANCE CRITERIA, DEMONSTRATING THAT THE LOT IS PERFORMING AS EXPECTED. DEVICE HISTORY REVIEW DID NOT IDENTIFY ANY ISSUES WITH THE COMPLAINT LOT. LABELING WAS REVIEWED AND WAS FOUND TO ADDRESS THE CUSTOMER REPORTED ISSUE. BASED ON THE AVAILABLE INFORMATION, NO SYSTEMIC ISSUE OR DEFICIENCY WAS IDENTIFIED FOR THE ARCHITECT FREE T4 REAGENT LOT 68900UD02 WAS IDENTIFIED.
THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 07K65-77 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 7K65, WITH 510K/PMA/BLA NUMBER K173122. AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
THE CUSTOMER REPORTED FALSELY ELEVATED ARCHITECT FREE T4 AND PROVIDED THE FOLLOWING DATA: INITIAL TEST RESULT WAS 24.87 PMOL/L, RE-TEST RESULT WAS 14.27 PMOL/L (CUSTOMER REFERENCE RANGE: 9.01-19.05). PER THE CUSTOMER DISCREPANT RESULTS WERE NOT REPORTED OUT TO THE MEDICAL PROVIDER. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
THE CUSTOMER REPORTED FALSELY ELEVATED ARCHITECT FREE T4 AND PROVIDED THE FOLLOWING DATA:INITIAL TEST RESULT WAS 24.87 PMOL/L, RE-TEST RESULT WAS 14.27 PMOL/L (CUSTOMER REFERENCE RANGE: 9.01-19.05).PER THE CUSTOMER DISCREPANT RESULTS WERE NOT REPORTED OUT TO THE MEDICAL PROVIDER. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1689664 | ARCHITECT FREE T4 REAGENT KIT | RADIOIMMUNOASSAY, FREE THYROXINE | CEC | ABBOTT IRELAND DIAGNOSTICS DIVISION | 68900UD02 | 00380740173678 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | ARC I2000SR INST, 03M74-02, (B)(6)| ARC I2000SR INST, 03M74-02, (B)(6) |