FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 2159651 · Received July 12, 2011

Report

Report Number
2050012-2011-02991
Event Type
Malfunction
Date Received
July 12, 2011
Date of Event
June 10, 2011
Report Date
June 10, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K060256
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A BEC FIELD SERVICE ENGINEER REPLACED VERTICAL HOME SENSOR AND MIXING PADDLES. FSE REALIGNED MIXER ASSEMBLY AND THIS RESOLVE THE RESOLVE.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) TO REPORT A LEAK ON THE CARTRIDGE CHEMISTRY (CC) REAGENT MIXER MOTION ERROR AND PUDDLE ON THE CC SIDE OF THE UNICEL DXC 600I INSTRUMENT. NO INJURY OR EXPOSURE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER FOR CLINICAL USE. JJE BECKMAN COULTER INC. N/A N/A

Patients

Seq Age Sex Outcome Treatment
1