CONTOUR®
Report
- Report Number
- 1810909-2025-00047
- Event Type
- Malfunction
- Date Received
- March 13, 2025
- Date of Event
- January 18, 2025
- Report Date
- March 13, 2025
- Manufacturer
- ASCENSIA DIABETES CARE US INC.
- Product Code
- NBW
- UDI-DI
- 05016003772702
- PMA / PMN Number
- K062058
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IZ
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE PICTURES PROVIDED BY THE DISTRIBUTOR WERE REVIEWED, AND IT WAS DETERMINED THAT THE CONTOUR® TEST STRIPS FROM LOT NUMBER DP3JJ3F02C HAD EXPIRATION DATES PRINTED AS 01-NOV-2024 AND 01-NOV-2026. THE EXPIRATION DATE OF 01-NOV-2024 WAS PRINTED BESIDE THE "EXPIRATION SYMBOL", WHILE THE EXPIRATION DATE OF 01-NOV-2026 WAS PRINTED BESIDE THE "EXPIRY DATE". THE UDI NUMBER ON THE PACKAGE WAS CORRECT: (B)(4). BASED ON THE TRACEABILITY RECORDS OF THE CONTOUR® TEST STRIPS WITH LOT NUMBER DP3JJ3F02C, THIS PRODUCT WAS DISTRIBUTED IN INDIA WITH THE CORRECT EXPIRATION DATE OF 31-AUG-2025 BY ASCENSIA. THE CONTOUR® TEST STRIPS FROM LOT NUMBER DP3JJ3F02C WITH INCORRECT EXPIRATION DATE WAS THEN SOLD IN IRAQ. THE CAUSE OF THE ISSUE IS ASSOCIATED WITH COUNTERFEIT PRODUCT. THERE WAS NO PATIENT INVOLVEMENT, THEREFORE, NO INFORMATION WAS CAPTURED IN SECTION A. THE CONTOUR® TEST STRIPS WITH SKU NUMBER 7727 AND LOT NUMBER DP3JJ3F02C HAS THE 510K NUMBER OF K062058 AND UDI NUMBER (B)(4). THE DEVICE WAS DISTRIBUTED OUTSIDE THE US, THEREFORE, NO GUDID INFORMATION EXISTS.
AN ASCENSIA DISTRIBUTOR FROM IRAQ REPORTED THAT DURING A FIELD VISIT, THEY IDENTIFIED COUNTERFEIT CONTOUR® TEST STRIPS IN A PACKAGE CONTAINING 100 TEST STRIPS. THE LOT NUMBER DP3JJ3F02C WAS PRINTED ON THE CARTON OF THE CONTOUR® TEST STRIPS WITH INCORRECT EXPIRATION DATE. ALL OTHER INFORMATION ON THE CARTON WAS CORRECT. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1700891 | CONTOUR® | BLOOD GLUCOSE TEST STRIPS | NBW | ASCENSIA DIABETES CARE US INC. | 7727 | 3JJ3F02C | 05016003772702 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |