FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK FLEXLINK

MDR report key: 21591754 · Received March 13, 2025

Report

Report Number
3003442380-2025-03585
Event Type
Malfunction
Date Received
March 13, 2025
Date of Event
November 20, 2024
Report Date
June 11, 2025
Manufacturer
UNOMEDICAL UM-D
Product Code
FPA
PMA / PMN Number
K143446
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

INITIAL AND FINAL MDR (B)(4). DEVICE 4 OF 4.

Additional Manufacturer Narrative · 0

SUPPLEMENTAL REPORT 01 - MDR (B)(4) - MDR 3003442380-2025-03585. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: THE INFORMATION IN THIS COMPLAINT PR (B)(4) HAS BEEN EVALUATED. THE BATCH 6001717 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. INVESTIGATION PROCESS OF THE COMPLAINT WAS CARRIED OUT IN ACCORDANCE WITH WORK INSTRUCTION (WI) GUIDANCE FOR VISUAL TESTING FOR COMPLAINTS AREA VERSION 3 FOR THE CODE SOFT CANNULA DISLODGED FROM TISSUE DURING USE. COMPLAINT INVESTIGATIONS. PHOTO/SAMPLE WAS NOT PROVIDED. IN ORDER TO TEST THE PRODUCT, THE REFERENCE SAMPLES FROM THE LOT HAVE BEEN REQUESTED. TEST RESULTS: VISUAL TEST ACCORDING TO WI VERSION 3 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6001717 WAS MANUFACTURED ACCORDING TO THE WI VERSION 44 AND PACKAGING IN THE MULTIVAC M07, ON 29/JUL/2023, WITH A TOTAL OF (B)(4) UNITS. ASSEMBLING OF FLEX SET CANNULA: THE LOT 3F00254 WAS MANUFACTURED ACCORDING TO THE WI VERSION 35 ASSEMBLED IN THE LINE L-02, ON 25/JUL/2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED RELATED TO THE MALFUNCTION REPORTED, NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 03/APR/2025 AGAINST MALFUNCTION CODE SOFT CANNULA DISLODGED FROM TISSUE DURING USE AND LOT 6001717 AND NO OTHER COMPLAINT HAS BEEN REGISTERED IN DATABASE FOR THE SAME LOT 6001717 AND MALFUNCTION CODE. CONCLUSION SUMMARY OF THE RELATED EVENT: AS A RESULT OF THE FOLLOWING: NO DEFECT ON THE VISUAL INSPECTION FOR REFERENCE SAMPLES, NO HARM REPORTED, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION, NO OTHER COMPLAINT RECEIVED ON THE LOT IN QUESTION AND REPORTED MALFUNCTION. NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CAPA PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Additional Manufacturer Narrative · 0

SUPPLEMENTAL REPORT 02 - MDR (B)(4). - MDR 3003442380-2025-03585. CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT THE SUBMITTED BRAND NAME UNDER D1, COMMON DEVICE NAME UNDER D2, MODEL NUMBER, SERIAL NUMBER AND PRIMARY UNIQUE DEVICE IDENTIFIER (UDI) NUMBER UNDER D4. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: D9: DEVICE AVAILABLE FOR EVALUATION HAS BEEN SELECTED AS YES & DATE RETURNED TO MFG WAS ADDED AS WELL. H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS.

Additional Manufacturer Narrative · 0

SUPPLEMENTAL REPORT 03 - (B)(4). CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT THE SUBMITTED PMA/510(K) NUMBER UNDER G4. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: THE INFORMATION IN THIS COMPLAINT (B)(4) HAS BEEN EVALUATED. THE BATCH 6001717 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. INVESTIGATION PROCESS OF THE COMPLAINT WAS CARRIED OUT IN ACCORDANCE WITH WORK INSTRUCTION (WI) GUIDANCE FOR VISUAL TESTING FOR COMPLAINTS AREA VERSION 3 FOR THE CODE SOFT CANNULA DISLODGED FROM TISSUE DURING USE. COMPLAINT INVESTIGATIONS. PHOTO/SAMPLE WAS NOT PROVIDED. IN ORDER TO TEST THE PRODUCT, THE REFERENCE SAMPLES FROM THE LOT HAVE BEEN REQUESTED. TEST RESULTS: VISUAL TEST ACCORDING TO WI VERSION 3 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6001717 WAS MANUFACTURED ACCORDING TO THE WI VERSION 44 AND PACKAGING IN THE MULTIVAC M07, ON 29/JUL/2023, WITH A TOTAL OF (B)(4) UNITS. ASSEMBLING OF FLEX SET CANNULA: THE LOT 3F00254 WAS MANUFACTURED ACCORDING TO THE WI VERSION 35 ASSEMBLED IN THE LINE L-02, ON 25/JUL/2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED RELATED TO THE MALFUNCTION REPORTED, NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 03/APR/2025 AGAINST MALFUNCTION CODE SOFT CANNULA DISLODGED FROM TISSUE DURING USE AND LOT 6001717 AND NO OTHER COMPLAINT HAS BEEN REGISTERED IN DATABASE FOR THE SAME LOT AND MALFUNCTION CODE. CONCLUSION SUMMARY OF THE RELATED EVENT: AS A RESULT OF THE FOLLOWING: NO DEFECT ON THE VISUAL INSPECTION FOR REFERENCE SAMPLES, NO HARM REPORTED, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION, NO OTHER COMPLAINT RECEIVED ON THE LOT IN QUESTION AND REPORTED MALFUNCTION. NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CAPA PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN BRAZIL. IT WAS REPORTED THAT THE PATIENT EXPERIENCED FOUR EVENTS WHERE THE CANNULA'S DETACHED FROM PATIENT'S SKIN. THE ISSUE OCCURED WITH FOUR INFUSION SETS ON 16-NOV-2024, 18-NOV-2024, 20-NOV-2024 AND 22-NOV-2024. THE SITE OF INSERTION WAS FLNAKS AND THIGHS. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

TO DATE, ADDITIONAL PATIENT OR EVENT DETAILS WERE RECEIVED WHICH HAVE BEEN ADDED IN H11.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
991196 ACCU-CHEK FLEXLINK UNO INSET I 12/6 GREY TCAP 10PK INT FPA UNOMEDICAL UM-D 06593941001 6001717

Patients

Seq Age Sex Outcome Treatment
1 7 YR Unknown