SMARTPHONE IOS APP: PUMP CONNECT
Report
- Report Number
- 2032227-2025-134654
- Event Type
- Malfunction
- Date Received
- March 13, 2025
- Date of Event
- February 12, 2025
- Report Date
- March 13, 2025
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- PKU
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- 003
Narratives
SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. AN ATTEMPT TO REPRODUCE THE REPORTED ISSUE WAS NOT PERFORMED AS IT WAS CONFIRMED THROUGH PREVIOUS ISSUE REFERENCES. THE SOFTWARE DID NOT SUCCESSFULLY ADHERE TO THE SPECIFIED REQUIREMENTS AND DID NOT PERFORM IN ACCORDANCE WITH THE EXPECTATIONS SPECIFIED IN THE (B)(4). AFTER CONDUCTING A THOROUGH INVESTIGATION, WE HAVE FOUND THAT THE REASON FOR THE REPORTED BEHAVIOR IS A KNOWN IOS DEFECT RESULTING IN A "DEVICE NOT FOUND" MESSAGE ON THE PUMP. THE ESF NUMBER THAT CORRESPONDS TO THE REPORTED ISSUE IS: (B)(4). TO ASSIST WITH THE RESOLUTION OF THE ISSUE, WE PROVIDED HELPLINE TEAM WITH THE FOLLOWING STEP TO ENSURE THAT IT IS ADDRESSED EFFECTIVELY: PLEASE TRY RESETTING THE BLUETOOTH CACHE AND RE-PAIRING WITH THE PUMP USING THE FOLLOWING STEPS: 1. LEAVE THE PUMP PAIRING SCREEN IN THE MINIMED MOBILE APP IF IT ACTIVE. 2. REMOVE THE PUMP FROM THE IOS BLUETOOTH SETTINGS (SETTINGS > BLUETOOTH > PUMP XXXXXXXX > FORGET THIS DEVICE). 3. REMOVE THE MOBILE DEVICE FROM THE PUMP. 4. TURN BLUETOOTH OFF FROM THE IOS SETTINGS APP (SETTINGS > BLUETOOTH). 5. WAIT 5 MINUTES. THIS WAIT IS NEEDED TO ALLOW THE IOS TO RESET THE BLUETOOTH CACHE. 6. TURN BLUETOOTH BACK ON. 7. NAVIGATE TO THE PUMP PAIRING SCREEN IN THE MINIMED MOBILE APP. 8. ATTEMPT PAIRING WITH THE PUMP IF THE PREVIOUS STEPS DID NOT WORK, PLEASE TRY RESTARTING YOUR MOBILE DEVICE AND RE-PAIRING WITH THE PUMP USING THE FOLLOWING STEPS: 1. LEAVE THE PUMP PAIRING SCREEN IN THE MINIMED MOBILE APP IF IT ACTIVE. 2. REMOVE THE PUMP FROM THE IOS BLUETOOTH SETTINGS (SETTINGS > BLUETOOTH > PUMP XXXXXXXX > FORGET THIS DEVICE). 3. REMOVE THE MOBILE DEVICE FROM THE PUMP. 4. RESTART THE MOBILE DEVICE. 5. NAVIGATE TO THE PUMP PAIRING SCREEN IN THE MINIMED MOBILE APP. 6. ATTEMPT PAIRING WITH THE PUMP IF THE PREVIOUS STEPS DID NOT WORK, PLEASE TRY RE-INSTALLING THE MMM APP USING THE FOLLOWING STEPS: 1. REMOVE THE MINIMED MOBILE APP AND ALL THE APP DATA (SETTINGS > GENERAL > IPHONE STORAGE > MINIMED MOBILE > DELETE APP). 2. REMOVE THE PUMP FROM THE IOS BLUETOOTH SETTINGS (SETTINGS > BLUETOOTH > PUMP XXXXXXXX > FORGET THIS DEVICE). 3. REMOVE THE MOBILE DEVICE FROM THE PUMP. 4. INSTALL THE MINIMED MOBILE APP. 5. NAVIGATE TO THE PUMP PAIRING SCREEN IN THE MINIMED MOBILE APP. 6. ATTEMPT PAIRING WITH THE PUMP. THE HELPLINE HAS NOT PROVIDED FEEDBACK ON THE RECOMMENDED STEPS AND WHETHER THE ISSUE HAS BEEN RESOLVED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER REPORTED A LOSS OF COMMUNICATION BETWEEN THE INSULIN PUMP AND MOBILE APPLICATION. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT MMT-6102. TROUBLESHOOTING WAS PERFORMED. UNABLE TO RESOLVE WITH EXISTING TROUBLESHOOTING OR LABELED INSTRUCTIONS, ISSUE ESCALATED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. NO PRODUCT RETURN IS REQUIRED FOR MMT-6102.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 804067 | SMARTPHONE IOS APP: PUMP CONNECT | INSULIN PUMP SECONDARY DISPLAY | PKU | MEDTRONIC MINIMED | MMT-6102 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 14 YR | Unknown |