FDA Adverse Event Malfunction Summary report: N

LOW PROFILE SCREW, 2.0 X 12MM, CORTICAL

MDR report key: 21582791 · Received March 12, 2025

Report

Report Number
1220246-2025-00884
Event Type
Malfunction
Date Received
March 12, 2025
Date of Event
February 21, 2025
Report Date
September 8, 2025
Manufacturer
ARTHREX, INC.
Product Code
HRS
UDI-DI
00888867311718
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED UPON AVAILABILITY OF ADDITIONAL INFORMATION.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: B5, D1, D4, H3, H6.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: G3, H3, H6. BASED ON THE INFORMATION PROVIDED WHICH MAY INCLUDE THE DEVICE (IF AVAILABLE AND RETURNED), PICTURES, VIDEOS, EVENT DESCRIPTION, AND ANY ADDITIONAL INFORMATION FROM THE FIELD, ARTHREX WAS ABLE TO CONCLUDE A MOST LIKELY CAUSE. THE MOST LIKELY CAUSE FOR THE REPORTED FAILURE CAN BE ATTRIBUTED TO USER ERROR OF THE DEVICE DUE TO EXCESSIVE FORCE BEING USED WHEN TENSIONING THE SCREW.

Description of Event or Problem · 0

ON (B)(6) 2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT TWO 2.0 MM SCREW HEADS SNAPPED OFF. THE FIRST SCREW HEAD BROKE OFF DURING INSERTION AND THE SECOND ONE BROKE WHEN IT WAS BEING REMOVED, IN EXCHANGE FOR A SHORTER SCREW. THE FRAGMENTS WERE REMOVED. THIS WAS DISCOVERED DURING A METACARPAL FRACTURE PROCEDURE ON (B)(6) 2025

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED ON 3/6/2025: THE PART NUMBER FOR THE SCREWS ARE AR-18720V-14 VAL SCREW AND AR-18720-12 CORTICAL SCREW. THE BODY OF BOTH SCREWS REMAINS IN THE PATIENT. THE CASE WAS COMPLETED USING AN AR-18720V-14 VAL SCREW AND AN AR-18720-12 CORTICAL SCREW. THE CASE WAS DELAYED FOR ABOUT FIVE MINUTES WITHOUT ADDITIONAL ANESTHESIA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
958729 LOW PROFILE SCREW, 2.0 X 12MM, CORTICAL BONE FIXATION SCREW/PLATE HRS ARTHREX, INC. LOW PROFILE SCREW, 2.0 X 12MM, CORTICAL UNK 00888867311718

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown