FDA Adverse Event Injury Summary report: N

MINIMED QUICK-SET

MDR report key: 21580485 · Received March 12, 2025

Report

Report Number
3003442380-2025-03499
Event Type
Injury
Date Received
March 12, 2025
Date of Event
February 6, 2025
Report Date
May 7, 2025
Manufacturer
UNOMEDICAL UM-D
Product Code
FPA
UDI-DI
05705244017450
PMA / PMN Number
K160648
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED FOR RECORD (B)(4) ON 16-APR-2025. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6003015 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 77 ON 06-SEP-2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 16-APR-2025 AGAINST HARM CODE NO MALFUNCTION BASED ON COMPLAINT INFORMATION, MALFUNCTION CODE NO MALFUNCTION DESCRIBE AND LOT 6003015 AND NO MORE COMPLAINT HAS BEEN REGISTERED IN DATABASE FOR THE SAME LOT, HARM CODE AND MALFUNCTION CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION, ONLY ONE COMPLAINT RECEIVED ON THE LOT IN QUESTION AND HARM CODE, THE TEST ON REFERENCE SAMPLES WERE NOT TESTED SINCE A NO MALFUNCTION RELATED TO THE INFUSION SET WAS REPORTED, THEREFORE NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Additional Manufacturer Narrative · 0

E1: PATIENT CITY - (B)(6), PATIENT COUNTRY - UNITED ARAB EMIRATES.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED ARAB EMIRATES. IT WAS REPORTED THAT THE PATIENT WENT TO THE EMERGENCY ROOM (ER) AND EVENTUALLY GOT HOSPITALIZED ON (B)(6) 2025 DUE TO HYPERGLYCEMIA. THE HIGHEST BLOOD GLUCOSE LEVEL WAS OVER 550 MG/DL AT THE TIME OF EVENT AND THE PATIENT GOT TREATED WITH INSULIN PEN. THE PATIENT WAS ADMITTED FOR MORE THAN FORTY EIGHT HOURS. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1399302 MINIMED QUICK-SET UNO QUICK-SET 60/6 SC1 MECA FPA UNOMEDICAL UM-D MMT-399A 6003015 05705244017450

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Hospitalization| R