ALLEN LIFT ASSIST BEACH CHAIR
Report
- Report Number
- 3010216206-2025-00002
- Event Type
- Malfunction
- Date Received
- March 11, 2025
- Date of Event
- February 24, 2025
- Report Date
- March 11, 2025
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- CCX
- UDI-DI
- 00615521004125
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
THE BEACH CHAIR SHOULDER POSITIONER IS DESIGNED TO POSITION AND SUPPORT THE PATIENT¿S HEAD AND TORSO IN A VARIETY OF SURGICAL PROCEDURES INCLUDING, BUT NOT LIMITED TO ORTHOPEDIC SURGERIES LIKE, ROTATOR CUFF, TOTAL SHOULDER, REVERSE TOTAL SHOULDER, SLAP REPAIR AND OTHER SHOULDER SURGERY. THESE DEVICES ARE INTENDED TO BE USED BY HEALTHCARE PROFESSIONALS WITHIN THE OPERATING ROOM SETTING. THE BEACH CHAIR WAS RECEIVED BY BAXTER. UPON INSPECTION, IT WAS DETERMINED THAT THE GAS SHOCK FAILED AND THE BACK BOARD WAS BROKE. BAXTER TECHNICIAN REPLACED THE SHOCK AND BACK BOARD. BEACH CHAIR NOW FUNCTIONS AS IT SHOULD. BASED ON THIS INFORMATION, NO FURTHER ACTIONS ARE REQUIRED AT THIS TIME. ALTHOUGH THERE WAS NO REPORTED INJURY WITH THIS EVENT, IF THE REPORT OF A BEACH CHAIR BACKBOARD BEING CRACKED WERE TO RECUR, IT COULD POTENTIALLY CAUSE SERIOUS INJURY OR DEATH. THEREFORE BAXTER IS REPORTING THIS EVENT.
BAXTER RECEIVED A REPORT STATING THAT ALLEN LIFT ASSIST BEACH CHAIR BACKBOARD WERE CRACKED AND WAS RECLINING SLIGHTLY. THE BED WAS LOCATED AT THE ACCOUNT AND OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, NO HARM OR SIGNIFICANT IMPACT TO THE SURGICAL PROCEDURE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1308938 | ALLEN LIFT ASSIST BEACH CHAIR | SUPPORT, PATIENT POSITION | CCX | BAXTER HEALTHCARE CORPORATION | A-91501 | N/A | 00615521004125 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |