FDA Adverse Event Injury Summary report: N

INJECTOR OPVSU SI NPS

MDR report key: 21567428 · Received March 10, 2025

Report

Report Number
1518293-2025-00002
Event Type
Injury
Date Received
March 10, 2025
Date of Event
February 6, 2025
Report Date
March 10, 2025
Manufacturer
LIEBEL-FLARSHEIM
Product Code
IZQ
UDI-DI
10746190002901
PMA / PMN Number
K063503
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

OVERALL INVESTIGATION SUMMARY SERVICE INVESTIGATED CUSTOMER REPORT OF AN AIR BUBBLE PROBLEM. SERVICE FOUND NO ALARM FAULTS, AND OBSERVED NO PROBLEMS DURING TESTING OF THE UNIT. THE USER WAS ADVISED TO DECREASE THE FILLING SPEED FROM MAX TO MIN. THE UNIT WAS TESTED FOR PERFORMANCE AND RETURNED TO SERVICE. INJECTOR COMPLIANT. A REVIEW OF CTS SHOWS NO SIMILAR ISSUE REPORTED ON THIS UNIT. IMPACT ASSESSMENT SUMMARY NO INJURY TO THE PATIENT/USER REPORTED IMDRF CODES: B01; C23; D11 ROOT / PROBABLE CAUSE CODE PROCESS/METHODS - INADEQUATE/INCORRECT PROCEDURE ROOT / PROBABLE CAUSE SUMMARY REFER TO INVESTIGATION SUMMARY. NO ADDITIONAL CAPA REQUIRED AT THIS TIME. GUERBET QUALITY WILL CONTINUE TO MONITOR AND TREND FOR SIMILAR ISSUES. THESE TRENDS AND ISSUES ARE REPORTED ON DURING QUALITY METRICS REVIEW AND DURING THE MANAGEMENT REVIEWS TO CONSIDER INPUT FOR ADDITIONAL CORRECTIVE ACTION. DISPOSITION SUMMARY UNIT REMAINED IN SERVICE.

Description of Event or Problem · 0

THIS CASE WAS REPORTED BY A FACILITY IN FRANCE ON 10 FEBRUARY 2025. CUSTOMER STATES THAT ON (B)(6) 2025 THAT AFTER THE INJECTION USING THE OPTIVANTAGE INJECTOR, AIR WAS INJECTED. INCIDENT AT THE IMAGERIE MEDICALE LEVALLOIS PERRET SCANNER CENTRE. DURING A COROSCANNER AT 15:59, AIR BUBBLES WERE INJECTED, DETECTED BY THE RADIOLOGIST. THESE BUBBLES DID NOT APPEAR IN THE CALCIUM SCORE (ACQUISITION OF THE CALCIUM SCORE THEN AUTOMATIC INJECTION FOR THE ACQUISITION OF THE HEART). THE PATIENT WAS PLACED ON A STRETCHER IN DORSAL DECUBITUS POSITION AND GIVEN OXYGEN AT 15 UNTIL THE ARRIVAL OF THE AMBULANCE. THE PATIENT'S BLOOD PRESSURE WAS GOOD (14/7) AND SHE WAS FEELING WELL. SHE WAS TRANSFERRED TO GARCHES AND PLACED IN A HYPERBARIC CHAMBER. ON ARRIVAL FOR EXAMINATION, THE PATIENT HAD A SLIGHTLY ELEVATED HEART RATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2031565 INJECTOR OPVSU SI NPS INJECTOR OPVSU SI NPS IZQ LIEBEL-FLARSHEIM 849001 10746190002901

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other