FDA Adverse Event Malfunction Summary report: N

COBAS 6000 E601 MODULE

MDR report key: 21560586 · Received March 10, 2025

Report

Report Number
1823260-2025-00682
Event Type
Malfunction
Date Received
March 10, 2025
Date of Event
February 11, 2025
Report Date
April 4, 2025
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JJE
PMA / PMN Number
K060373
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE TROPONIN T STAT REAGENT LOT NUMBER WAS 802253 WITH AN EXPIRATION DATE OF 31-OCT-2025.

Additional Manufacturer Narrative · 0

QC WAS ACCEPTABLE. THERE IS NO INDICATION OF A REAGENT PERFORMANCE ISSUE. NO ISSUES WERE IDENTIFIED DURING A REVIEW OF THE ALARM TRACE DATA. THE FIELD SERVICE ENGINEER (FSE) FOUND THE WATER PRESSURE WAS TOO HIGH. THE FSE ADJUSTED THE WATER PRESSURE AND THE GEAR PUMP PRESSURE. THE SERVICE MAINTENANCE ACTIONS RESOLVED THE ISSUE.

Description of Event or Problem · 0

THE INITIAL REPORTER QUESTIONED HIGH RESULTS FOR 2 PATIENT SAMPLES TESTED FOR ELECSYS TROPONIN T GEN 5 STAT (TROPONIN T STAT) ON A COBAS 6000 E601 MODULE. DISCREPANT RESULTS WERE PROVIDED FOR 1 PATIENT SAMPLE. THE INITIAL RESULT WAS 48.07 PG/ML. THIS RESULT WAS REPORTED OUTSIDE OF THE LABORATORY WHERE THE DOCTOR QUESTIONED IT. THE REPEAT RESULT FROM A DIFFERENT E601 MODULE WAS 18.10 PG/ML. THIS RESULT WAS BELIEVED TO BE CORRECT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2144486 COBAS 6000 E601 MODULE IMMUNOCHEMISTRY ANALYZER JJE ROCHE DIAGNOSTICS E601

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown