ARCTIC SUN® 5000
Report
- Report Number
- 1018233-2025-01567
- Event Type
- Malfunction
- Date Received
- March 10, 2025
- Date of Event
- March 5, 2025
- Report Date
- April 30, 2025
- Manufacturer
- MEDIVANCE, INC. ¿ 1725056
- Product Code
- DWJ
- UDI-DI
- 00801741127755
- PMA / PMN Number
- K161602
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. UDI FORMAT UPDATED WITH AVAILABLE PRODUCT INFORMATION PER GUDID. H11: SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD.
THE REPORTED ISSUE WAS CONFIRMED. THE ROOT CAUSE IDENTIFIED IS A WEAK CIRCULATION PUMP. THE DEVICE WAS EVALUATED UPON RECEIPT. PRIMED TO FILL. SERVICED CIRCULATION PUMP. CTU FLOW IS READING 500 ML/M LOW DUE TO AN IMPAIRED IMPELLER. THE CIRCULATION AND MIXING PUMP MOTORS ARE MAKING A CLACKING SOUND WHEN COLD STARTED. PERFORMED PM PROCEDURE. SERVICED MIXING PUMP. REPLACED HEATER, MANIFOLD O-RINGS, TANK TUBING, CIRCULATION PUMP MOTOR, MIXING PUMP MOTOR, FLOWMETER. THE ARCTIC SUN 5000 PASSED ALL PERFORMANCE TESTING, CALIBRATION AND ELECTRICAL SAFETY TESTS AND IS FUNCTIONING PROPERLY AND READY FOR USE. A DHR IS NOT REQUIRED AS THE REPORTED EVENT IS NOT AN OUT OF BOX FAILURE AND THEREFORE THE REPORTED EVENT IS NOT MANUFACTURING RELATED. THE COMPLAINT OR REPORTED ISSUE WAS CONFIRMED THROUGH OTHER ELEMENTS OF THE INVESTIGATION TO NOT BE LABELING OR PACKAGING RELATED. UDI FORMAT UPDATED WITH AVAILABLE PRODUCT INFORMATION PER GUDID. H11: SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD.
IT WAS REPORTED THAT THE ARCTIC SUN DEVICE WOULD NOT FILL. NO SUCTION AT LEFT PORT, FAILED CIRCULATION PUMP.
IT WAS REPORTED THAT THE ARCTIC SUN DEVICE WOULD NOT FILL. NO SUCTION AT LEFT PORT, FAILED CIRCULATION PUMP. PER SAMPLE EVALUATION RESULTS ON 15APR2025, CTU FLOW WAS READING 500 ML/M LOW DUE TO AN IMPAIRED IMPELLER. THE CIRCULATION PUMP MOTOR WAS MAKING A CLACKING SOUND WHEN COLD STARTED. THE MIXING PUMP MOTOR WAS MAKING A CLACKING SOUND WHEN COLD STARTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1197259 | ARCTIC SUN® 5000 | ARCTIC SUN DEVICE | DWJ | MEDIVANCE, INC. ¿ 1725056 | NA | 00801741127755 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Other |