SIGMA SPECTRUM INFUSION PUMP
Report
- Report Number
- 1314492-2025-00513
- Event Type
- Malfunction
- Date Received
- March 10, 2025
- Report Date
- March 10, 2025
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- UDI-DI
- 00085412498683
- PMA / PMN Number
- K133801
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS RECEIVED FOR EVALUATION. DURING EVALUATION, THE REPORTED PROBLEM OF '3 BUTTONS WERE NOT RESPONDING ON A SPECTRUM PUMP, INCLUDING THE DECIMAL BUTTON, NUMBER 4 (ALSO ¿JKL¿ BUTTON) AND NUMBER 5 (ALSO ¿MNO¿ BUTTON)' WAS REPRODUCED. A SERVICE HISTORY REVIEW REVEALED NO INDICATION THAT THE PARTS REPLACED DURING SERVICING CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE CONDITION WAS DETERMINED TO BE A FAULTY KEYPAD. THE KEYPAD WILL BE REPLACED TO CORRECT THE REPORTED PROBLEM. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT 3 BUTTONS WERE NOT RESPONDING ON A SPECTRUM PUMP, INCLUDING THE DECIMAL BUTTON, NUMBER 4 (ALSO ¿JKL¿ BUTTON) AND NUMBER 5 (ALSO ¿MNO¿ BUTTON). THIS OCCURRED DURING TESTING IN THE BIOMED SERVICE DEPARTMENT. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1450864 | SIGMA SPECTRUM INFUSION PUMP | PUMP, INFUSION | FRN | BAXTER HEALTHCARE CORPORATION | 35700BAX2 | NA | 00085412498683 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |